A dynamic reconstruction and motion estimation framework for cardiorespiratory motion-resolved real-time volumetric MR imaging (DREME-MR)
Based on a 3D pre-treatment magnetic resonance (MR) scan, we developed DREME-MR to jointly reconstruct the reference patient anatomy and solve a data-driven, patient-specific cardiorespiratory motion model. Via a motion encoder simultaneously learned during the reconstruction, DREME-MR further enables real-time volumetric MR imaging and cardiorespiratory motion tracking with minimal intra-treatment k-space data.

Approach: DREME-MR integrates dynamic MRI reconstruction and real-time MR imaging into a unified, dual-task learning framework. From a 3D radial-spoke-based pre-treatment MR scan, DREME-MR uses spatiotemporal implicit-neural-representation (INR) to reconstruct pre-treatment dynamic volumetric MR images (learning task 1). The INR-based reconstruction takes a joint image reconstruction and deformable registration approach, yielding a reference anatomy and a corresponding cardiorespiratory motion model. The motion model adopts a low-rank, multi-resolution representation to approximate motion fields as products of motion coefficients and motion basis components (MBCs). Via a progressive, frequency-guided strategy, DREME-MR decouples cardiac MBCs from respiratory MBCs to resolve the two distinct motion modes. Simultaneously with the pre-treatment dynamic MRI reconstruction, DREME-MR also trains a multilayer perceptron (MLP)-based motion encoder to infer cardiorespiratory motion coefficients directly from the raw k-space data (learning task 2), allowing real-time, intra-treatment volumetric MR imaging and motion tracking with minimal k-space data (20-30 spokes) acquired after the pre-treatment MRI scan.

Main results: Evaluated using data from a digital phantom (XCAT) and a human scan, DREME-MR solves real-time 3D cardiorespiratory motion with a latency of < 165 ms (= 150-ms data acquisition + 15-ms inference time), fulfilling the temporal constraint of real-time imaging. The XCAT study achieves mean(±S.D.) center-of-mass tracking errors of 0.73±0.38mm for a lung tumor and 1.69±1.12mm for the left ventricle. The human study shows good motion correlations (liver: 0.96; left ventricle: 0.65) between DREME-MR-solved motion and extracted surrogate signals.

Significance: DREME-MR allows real-time 3D MRI and cardiorespiratory motion tracking with low latency, advancing intra-treatment MR-guided adaptive radiotherapy, including real-time multileaf collimator (MLC) tracking.
Authors
- Jie Deng (ORCID: https://orcid.org/0000-0003-0249-1969)
- Xiaoxue Qian (ORCID: https://orcid.org/0000-0002-6185-1074)
- Ricardo Otazo (ORCID: https://orcid.org/0000-0002-3782-4930)
- You Zhang (ORCID: https://orcid.org/0000-0002-8033-2755)
- Guoping Xu (ORCID: https://orcid.org/0000-0001-8932-3321)
- Hua‐Chieh Shao (ORCID: https://orcid.org/0009-0003-7664-7257)
- Can Wu (ORCID: https://orcid.org/0000-0003-1171-4546)
Institutions
- Memorial Sloan Kettering Cancer Center (US)
- Southwestern Medical Center (US)
- Artificial Intelligence in Medicine (Canada) (CA)
- Southwestern Medical Center (US)
- The University of Texas Southwestern Medical Center (US)
Publication Details
- Journal
- Physics in Medicine and Biology
- Published
- 2025-08-08
- DOI
- https://doi.org/10.1088/1361-6560/adf9b9
- Citations
- 1
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 1.36
Funders
- Varian Medical Systems
- National Institutes of Health