Evaluation of corneal endothelium damage with a new handheld fragmentation device

The overall differences among the 3 groups regarding intact, damaged, and lost cells were statistically significant. However, by applying the Bonferroni correction, only the post-hoc pairwise comparisons between groups 1 (phaco tip) and 2 (miCOR tip, half throttle, without lens agitator engaged) were statistically significant regarding the 3 above-mentioned parameters.Conclusion: miCOR is a handheld, finger-controlled lens removal device with a blunt and rounded tip that vibrates at a low frequency, using mechanical agitation for nuclear disassembly. This acute rabbit study suggests CEC damage from the device, when the lens agitator is engaged is comparable to that of standard torsional phacoemulsification.

Authors

Institutions

Publication Details

Journal
Journal of Cataract & Refractive Surgery
Published
2025-04-09
DOI
https://doi.org/10.1097/j.jcrs.0000000000001668
Primary Topic
Intraocular Surgery and Lenses
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of corneal endothelium damage with a new handheld fragmentation device

Kevin Eid, Jeff H. Pettey, Neil Kelkar, Emilie L. Ungricht et al.
Journal of Cataract & Refractive Surgery
Intraocular Surgery and Lenses
article

Evaluation of corneal endothelium damage with a new handheld fragmentation device

Kevin Eid, Jeff H. Pettey, Neil Kelkar, Emilie L. Ungricht, Liliana Werner, Reiker G Ricks
article en

Abstract

The overall differences among the 3 groups regarding intact, damaged, and lost cells were statistically significant. However, by applying the Bonferroni correction, only the post-hoc pairwise comparisons between groups 1 (phaco tip) and 2 (miCOR tip, half throttle, without lens agitator engaged) were statistically significant regarding the 3 above-mentioned parameters.Conclusion: miCOR is a handheld, finger-controlled lens removal device with a blunt and rounded tip that vibrates at a low frequency, using mechanical agitation for nuclear disassembly. This acute rabbit study suggests CEC damage from the device, when the lens agitator is engaged is comparable to that of standard torsional phacoemulsification.

Journal of Cataract & Refractive SurgeryVol. 51(7)
Intermountain Healthcare (US)
Openalex Percentile: Top 88%
Intraocular Surgery and Lenses
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Evaluation of corneal endothelium damage with a new handheld fragmentation device — Kevin Eid, Jeff H. Pettey, et al. · Journal of Cataract & Refractive Surgery (2025) | TGRS Research Map | TGRS