Liquid chromatography tandem mass spectrometry (LC-MS/MS) candidate reference measurement procedure for urine albumin
This RMP demonstrated excellent imprecision, achieving an overall CV of 2.8 % and meeting the CV ≤6.2 % performance specification required for standardizing urine albumin measurements in clinical laboratories.
Authors
- Michael D. Evans (ORCID: https://orcid.org/0000-0001-7449-3993)
- Jesse C. Seegmiller (ORCID: https://orcid.org/0000-0003-2542-6502)
- S. Shiba (ORCID: https://orcid.org/0009-0000-0968-5083)
- Zoi E. Sychev (ORCID: https://orcid.org/0000-0002-9830-5245)
Institutions
- University of Minnesota (US)
- Southern California Clinical and Translational Science Institute (US)
Publication Details
- Journal
- Clinical Chemistry and Laboratory Medicine (CCLM)
- Published
- 2025-04-08
- DOI
- https://doi.org/10.1515/cclm-2024-1374
- Citations
- 1
- Primary Topic
- Chronic Kidney Disease and Diabetes
- Type
- article
- Field-Weighted Citation Impact
- 1.43