First experimental demonstration of charge-cloud imaging for micrometer-scale resolution with a photon-counting silicon CT detector
Evaluation of a new sensor for micrometer-resolution photon-counting CT. DAC-sweeps are performed using a commercial x-ray tube and are compared to simulations. An edge-scan using a 250μm tungsten wafer without any interaction logic is also performed, as well as single interaction readout of the energy spectrum that is compared to simulations. The edge-scan shows a line spread function with a full width at half maximum of 11.6μm and a 5% modulation transfer function at 850 lp/cm. Fair agreement with simulations indicated that employing the interaction can further significantly improve spatial resolution.
Authors
- Mats Danielsson (ORCID: https://orcid.org/0000-0002-3039-9791)
- Mats Persson (ORCID: https://orcid.org/0000-0002-5092-8822)
- Rickard Brunskog (ORCID: https://orcid.org/0000-0002-3309-9276)
Institutions
- Karolinska University Hospital (SE)
- KTH Royal Institute of Technology (SE)
Publication Details
- Published
- 2025-04-08
- DOI
- https://doi.org/10.1117/12.3048609
- Primary Topic
- Advanced X-ray and CT Imaging
- Type
- article
- Field-Weighted Citation Impact
- 0.00