A Comparative Analysis of Glycemic Metrics Derived From Three Continuous Glucose Monitoring Systems
The CGM systems indicated discordant glycemic metrics, which should be considered in diabetes therapy. Different CGM systems should provide the same glucose readings and CGM-derived metrics when used by the same person.