Log file‐based quality assurance method for respiratory gating system
Log file analysis revealed precise temporal synchronization, with mean time differences of 0.03 s ± 0.05 s between the planned 25% phase and beam-on, and -0.04 s ± 0.05 s between 75% phase and beam-off. The log-derived beam-on duration (2.61 s ± 0.02 s) closely matched the planned duration (2.66 s ± 0.00 s). Three-month log data showed consistent temporal accuracy, with trigger-on times remaining stable at 2.60 s ± 0.01 s across all measurements. Supporting ion chamber measurements confirmed dosimetric agreement between gating and non-gating modes (difference: 0.05 cGy ± 0.09 cGy) CONCLUSIONS: The proposed log file-based QA method demonstrated high accuracy and reproducibility in assessing respiratory gating performance. This approach provides an efficient, objective method for standardizing QA procedures in respiratory-gated radiation therapy, enhancing treatment accuracy and patient safety.
Authors
- Wonjoong Cheon (ORCID: https://orcid.org/0000-0001-5721-3347)
- Kyu Hye Choi (ORCID: https://orcid.org/0000-0001-6446-6377)
- Yunji Seol
- Young Kyu Lee
- Hong Qi Tan (ORCID: https://orcid.org/0000-0001-7878-4544)
- C.S. Park
- Young‐Nam Kang
- Byung Ock Choi
Institutions
- Nanyang Technological University (SG)
- Duke-NUS Medical School (SG)
- National Cancer Centre Singapore (SG)
- The Catholic University of Korea Seoul St. Mary's Hospital (KR)
- Catholic University of Korea (KR)
Publication Details
- Journal
- Journal of Applied Clinical Medical Physics
- Published
- 2025-04-08
- DOI
- https://doi.org/10.1002/acm2.70101
- Primary Topic
- Advanced Radiotherapy Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Catholic University of Korea
- Korea Electrotechnology Research Institute
- National Research Council of Science and Technology
- Ministry of Science and ICT, South Korea