Performance Comparison of Oximetry Devices During 6-Minute Walk Tests

<b>Background:</b> Pulse oximetry is a common technique used to measure S<sub>pO<sub>2</sub></sub>, with increasing numbers of portable devices of different designs and technical specifications. This study examined accuracy of commercial and medical-grade portable oximetry devices (BioBeat wristwatch, Nonin WristOx<sub>2</sub> 3150, and Heart Sure A320) in measuring S<sub>pO<sub>2</sub></sub> compared with the reference standard (Masimo Rad-5) during 6-min walk tests (6MWTs). <b>Methods:</b> This prospective study included 50 subjects, with 10 recruited for each of the five target groups, including COPD, interstitial lung disease, pulmonary hypertension, oxygen therapy users, and other cardiac/respiratory diseases. All subjects underwent 6MWT using the reference standard and the three test oximetry devices. Accuracy was measured by comparing mean nadir S<sub>pO<sub>2</sub></sub>. Subjects were asked to rate their preference for different oximetry devices at test completion. <b>Results:</b> In contrast to the Masimo, the Nonin had lower nadir and mean S<sub>pO<sub>2</sub></sub> <i>(P</i> ≤ .001 for both), with higher measurements for the BioBeat (<i>P</i> < .001 for both) and no differences for the Heart Sure (<i>P</i> = .18 for nadir S<sub>pO<sub>2</sub></sub>, <i>P</i> <b>=</b> .96 for mean S<sub>pO<sub>2</sub></sub>). Time spent at S<sub>pO<sub>2</sub></sub> ≤ 88 was higher for the Nonin (70%, <i>P</i> < .001) but lower for the Heart Sure (58%, <i>P</i> < .001), compared with the Masimo (64%). There were no significant differences observed in peak heart rate measurements between the Masimo with the Nonin and BioBeat (<i>P</i> < .05 for both), with lower measurements using the Heart Sure (<i>P</i> < .001). The BioBeat was most preferable amongst subjects (<i>n =</i> 31), with all test oximeters having high patient satisfaction (mean \7-point Likert scale scores of 6-7). <b>Conclusion:</b> The Heart Sure A320 and Nonin Wrist-Ox<sub>2</sub> 3150 had comparable performance in detecting nadir S<sub>pO<sub>2</sub></sub> within a mean difference of 3% in comparison with the Masimo Rad-5, with potential use for remote monitoring.

Authors

Institutions

Publication Details

Journal
Respiratory Care
Published
2025-04-12
DOI
https://doi.org/10.1089/respcare.12259
Primary Topic
Chronic Obstructive Pulmonary Disease (COPD) Research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Performance Comparison of Oximetry Devices During 6-Minute Walk Tests

M. Horrigan, T Churchward, Karla Logie, Max Schulte et al.
Respiratory Care
Chronic Obstructive Pulmonary Disease (COPD) Research
article

Performance Comparison of Oximetry Devices During 6-Minute Walk Tests

M. Horrigan, T Churchward, Karla Logie, Max Schulte, Leona Dowman, Julian Gin, Yet Hong Khor, E. Marshall, Angeline Filiex
article en

Abstract

<b>Background:</b> Pulse oximetry is a common technique used to measure S<sub>pO<sub>2</sub></sub>, with increasing numbers of portable devices of different designs and technical specifications. This study examined accuracy of commercial and medical-grade portable oximetry devices (BioBeat wristwatch, Nonin WristOx<sub>2</sub> 3150, and Heart Sure A320) in measuring S<sub>pO<sub>2</sub></sub> compared with the reference standard (Masimo Rad-5) during 6-min walk tests (6MWTs). <b>Methods:</b> This prospective study included 50 subjects, with 10 recruited for each of the five target groups, including COPD, interstitial lung disease, pulmonary hypertension, oxygen therapy users, and other cardiac/respiratory diseases. All subjects underwent 6MWT using the reference standard and the three test oximetry devices. Accuracy was measured by comparing mean nadir S<sub>pO<sub>2</sub></sub>. Subjects were asked to rate their preference for different oximetry devices at test completion. <b>Results:</b> In contrast to the Masimo, the Nonin had lower nadir and mean S<sub>pO<sub>2</sub></sub> <i>(P</i> ≤ .001 for both), with higher measurements for the BioBeat (<i>P</i> < .001 for both) and no differences for the Heart Sure (<i>P</i> = .18 for nadir S<sub>pO<sub>2</sub></sub>, <i>P</i> <b>=</b> .96 for mean S<sub>pO<sub>2</sub></sub>). Time spent at S<sub>pO<sub>2</sub></sub> ≤ 88 was higher for the Nonin (70%, <i>P</i> < .001) but lower for the Heart Sure (58%, <i>P</i> < .001), compared with the Masimo (64%). There were no significant differences observed in peak heart rate measurements between the Masimo with the Nonin and BioBeat (<i>P</i> < .05 for both), with lower measurements using the Heart Sure (<i>P</i> < .001). The BioBeat was most preferable amongst subjects (<i>n =</i> 31), with all test oximeters having high patient satisfaction (mean \7-point Likert scale scores of 6-7). <b>Conclusion:</b> The Heart Sure A320 and Nonin Wrist-Ox<sub>2</sub> 3150 had comparable performance in detecting nadir S<sub>pO<sub>2</sub></sub> within a mean difference of 3% in comparison with the Masimo Rad-5, with potential use for remote monitoring.

Respiratory CareVol. 70(5)
The University of Melbourne (AU), Monash Alfred Psychiatry Research centre (AU), Institute for Breathing and Sleep (AU), Respiratory Clinical Trials (AU), Austin Health (AU)
Good health and well-being
Openalex Percentile: Top 89%
Chronic Obstructive Pulmonary Disease (COPD) Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.