Concentration Measurement with Ultrabroad Dynamic Range Using Few-Step Variable Optical-Path-Length Slope Method

Concentration measurement has important applications in many fields, including pollution assessment in environmental science and drug dosage calculation in biomedical research. In the conventional methods, concentration is determined by measuring absorbance along a fixed long optical path. However, it is not suitable for high-concentration measurement. Herein, we have proposed a few-step variable optical-path-length slope method (fs-VOSM) for ultrabroad dynamic-range concentration measurement. As a proof of the method, we devised an fs-VOSM system in which a reference path is included to enhance the accuracy and repeatability. The measurement is conducted at 5 positions along ultrashort optical path (0-20 μm) for 800 ms. In the measurement of potassium dichromate solution concentration, the fs-VOSM system exhibits a wide dynamic range from 0.879 to 70.726 g/L with coefficient of variation (CV < 1.4%) and high accuracy (relative error within ±3.5%). We prospect that the fs-VOSM can be widely adopted in many advanced instruments such as process analyzer, flow injection analyzer, and turbidity meter.

Authors

Institutions

Publication Details

Journal
Analytical Chemistry
Published
2025-04-07
DOI
https://doi.org/10.1021/acs.analchem.4c06129
Primary Topic
Analytical Chemistry and Sensors
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Concentration Measurement with Ultrabroad Dynamic Range Using Few-Step Variable Optical-Path-Length Slope Method

Xujun Yuan, Bo Dai, Songlin Zhuang, Wei Li et al.
Analytical Chemistry
Analytical Chemistry and Sensors
article

Concentration Measurement with Ultrabroad Dynamic Range Using Few-Step Variable Optical-Path-Length Slope Method

Xujun Yuan, Bo Dai, Songlin Zhuang, Wei Li, Dawei Zhang, Mingwei Zhao
article en

Abstract

Concentration measurement has important applications in many fields, including pollution assessment in environmental science and drug dosage calculation in biomedical research. In the conventional methods, concentration is determined by measuring absorbance along a fixed long optical path. However, it is not suitable for high-concentration measurement. Herein, we have proposed a few-step variable optical-path-length slope method (fs-VOSM) for ultrabroad dynamic-range concentration measurement. As a proof of the method, we devised an fs-VOSM system in which a reference path is included to enhance the accuracy and repeatability. The measurement is conducted at 5 positions along ultrashort optical path (0-20 μm) for 800 ms. In the measurement of potassium dichromate solution concentration, the fs-VOSM system exhibits a wide dynamic range from 0.879 to 70.726 g/L with coefficient of variation (CV < 1.4%) and high accuracy (relative error within ±3.5%). We prospect that the fs-VOSM can be widely adopted in many advanced instruments such as process analyzer, flow injection analyzer, and turbidity meter.

Analytical ChemistryVol. 97(16)
University of Shanghai for Science and Technology (CN)
National Natural Science Foundation of China
Climate action
Openalex Percentile: Top 90%
Analytical Chemistry and Sensors
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Concentration Measurement with Ultrabroad Dynamic Range Using Few-Step Variable Optical-Path-Length Slope Method — Xujun Yuan, Bo Dai, et al. · Analytical Chemistry (2025) | TGRS Research Map | TGRS