Evaluation of Image Quality of Temporal Maximum Intensity Projection and Average Intensity Projection of Adaptive 4D-Spiral CT Scans: A Phantom Study

Adaptive four-dimensional (4D) spiral computed tomography (CT) scans facilitate the acquisition of volume perfusion data for organs or long-range vessels; however, optimizing image quality and reducing noise while minimizing radiation doses remains challenging. Thus, image-processing techniques such as temporal maximum intensity projection (MIP) and average intensity projection (AIP) are crucial in this context. This ex vivo study aimed to compare the image noise, spatial resolution, and measurements of temporal MIP and AIP images generated from low radiation dose 4D CT scans data with those of conventional CT images using phantoms. Three phantoms were scanned with equivalent radiation doses using single helical and adaptive 10-phase 4D spiral scans using a third-generation dual-source CT scanner. Temporal MIP and AIP images of 4D CT scans were generated by summing varying numbers of phases, incorporating automatic motion correction with non-rigid registration and noise reduction algorithm. The CT values and image noise of the temporal MIP and AIP images were compared to conventional CT images. The task transfer function (TTF) was calculated using static phantoms. Vessel diameters of the phantoms for each image dataset were evaluated using motion phantoms. Temporal AIP images showed comparable CT values with those of the reference image. In contrast, the CT values of the temporal MIP images were significantly higher than those of the reference images (p<0.01). The image noise of temporal AIP images with six or more phases was equal to or lower than that of the reference images. In contrast, temporal MIP images exhibited consistently high noise levels regardless of the number of summed phases. The TTF of temporal AIP images was comparable to that of the reference CT images. However, the TTF of temporal MIP images gradually decreased as the number of summed phases increased. No significant differences were observed in vessel diameter measurements among the three groups or with varying numbers of summed phases (p>0.05). In conclusion, temporal MIP and AIP images generated from low radiation dose 4D CT scans could effectively reduce noise while preserving measurement reliability in the motion phantom, achieving performance comparable to conventional CT images.

Authors

Institutions

Publication Details

Journal
Cureus
Published
2025-04-07
DOI
https://doi.org/10.7759/cureus.81849
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of Image Quality of Temporal Maximum Intensity Projection and Average Intensity Projection of Adaptive 4D-Spiral CT Scans: A Phantom Study

Hiroki Horinouchi, Tatsuya Nishii, Noriyuki Negi, Toshinori Sekitani et al.
Cureus
Advanced X-ray and CT Imaging
article

Evaluation of Image Quality of Temporal Maximum Intensity Projection and Average Intensity Projection of Adaptive 4D-Spiral CT Scans: A Phantom Study

Hiroki Horinouchi, Tatsuya Nishii, Noriyuki Negi, Toshinori Sekitani, Keitaro Sofue, Satoru Takahashi, Tetsuya Fukuda
article en

Abstract

Adaptive four-dimensional (4D) spiral computed tomography (CT) scans facilitate the acquisition of volume perfusion data for organs or long-range vessels; however, optimizing image quality and reducing noise while minimizing radiation doses remains challenging. Thus, image-processing techniques such as temporal maximum intensity projection (MIP) and average intensity projection (AIP) are crucial in this context. This ex vivo study aimed to compare the image noise, spatial resolution, and measurements of temporal MIP and AIP images generated from low radiation dose 4D CT scans data with those of conventional CT images using phantoms. Three phantoms were scanned with equivalent radiation doses using single helical and adaptive 10-phase 4D spiral scans using a third-generation dual-source CT scanner. Temporal MIP and AIP images of 4D CT scans were generated by summing varying numbers of phases, incorporating automatic motion correction with non-rigid registration and noise reduction algorithm. The CT values and image noise of the temporal MIP and AIP images were compared to conventional CT images. The task transfer function (TTF) was calculated using static phantoms. Vessel diameters of the phantoms for each image dataset were evaluated using motion phantoms. Temporal AIP images showed comparable CT values with those of the reference image. In contrast, the CT values of the temporal MIP images were significantly higher than those of the reference images (p<0.01). The image noise of temporal AIP images with six or more phases was equal to or lower than that of the reference images. In contrast, temporal MIP images exhibited consistently high noise levels regardless of the number of summed phases. The TTF of temporal AIP images was comparable to that of the reference CT images. However, the TTF of temporal MIP images gradually decreased as the number of summed phases increased. No significant differences were observed in vessel diameter measurements among the three groups or with varying numbers of summed phases (p>0.05). In conclusion, temporal MIP and AIP images generated from low radiation dose 4D CT scans could effectively reduce noise while preserving measurement reliability in the motion phantom, achieving performance comparable to conventional CT images.

CureusVol. 17(4)
Kobe University Hospital (JP), National Cerebral and Cardiovascular Center (JP), Takatsuki General Hospital (JP), Kobe University (JP), Osaka Institute of Technology (JP)
Climate action
Openalex Percentile: Top 93%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.