Structural Updates to the Implant and Refill Needle of the Port Delivery Platform
Structural updates to the PD-P implant and refill needle resulted in the ability to withstand the equivalent of over 50 years of simulated use without septum dislodgement and should improve the longevity of the device in clinical use.
Authors
- Carlos Quezada-Ruiz (ORCID: https://orcid.org/0000-0002-9909-6553)
- Margaret Chang
- Ian Pearce (ORCID: https://orcid.org/0000-0002-7710-0531)
- Shamika Gune (ORCID: https://orcid.org/0000-0001-9963-5652)
- Natasha Singh (ORCID: https://orcid.org/0000-0002-5130-9679)
- Nancy M. Holekamp (ORCID: https://orcid.org/0000-0001-7850-8515)
- Joshua Horvath
- Shrirang V. Ranade (ORCID: https://orcid.org/0000-0002-9457-2477)
- Dante J. Pieramici
- Carl D. Regillo (ORCID: https://orcid.org/0000-0001-8260-9925)
- Peter A. Campochiaro
- Geoffrey Wise
Institutions
- Roche (Switzerland) (CH)
- Johns Hopkins University (US)
- University of Liverpool (GB)
- Thomas Jefferson University (US)
- Johns Hopkins Medicine (US)
- Royal Liverpool University Hospital (GB)
- California Retina Consultants (US)
- Retina Research Foundation (US)
- Wills Eye Hospital (US)
Publication Details
- Journal
- Translational Vision Science & Technology
- Published
- 2025-04-07
- DOI
- https://doi.org/10.1167/tvst.14.4.8
- Citations
- 3
- Primary Topic
- Maritime Navigation and Safety
- Type
- article
- Field-Weighted Citation Impact
- 3.44
Funders
- F. Hoffmann-La Roche