Key comparison BIPM.RI(I)-K8 of high dose-rate <sup>192</sup>Ir brachytherapy standards for reference air kerma rate of the NIM and the BIPM
Main text An indirect comparison of the standards for reference air kerma rate for 192 Ir high dose rate (HDR) brachytherapy sources of the National Institute of Metrology (NIM), China, and of the Bureau International des Poids et Mesures (BIPM) was carried out at the NIM in October 2024. The comparison result, based on the calibration coefficients for a transfer standard and expressed as a ratio of the NIM and the BIPM standards for reference air kerma rate, is 1.0027 with a combined standard uncertainty of 0.0045. To reach the main text of this paper, click on Final Report . Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/ . The final report has been peer-reviewed and approved for publication by the CCRI, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).
Authors
- Kun Wang (ORCID: https://orcid.org/0000-0002-9533-7452)
- C Kessler
- Ji Wang (ORCID: https://orcid.org/0000-0002-0724-7538)
- Peiwei Wang (ORCID: https://orcid.org/0000-0002-9712-8849)
- Han Jinpeng
Institutions
- National Institute of Metrology (CN)
- Bureau international des poids et mesures (FR)
Publication Details
- Journal
- Metrologia
- Published
- 2025-01-01
- DOI
- https://doi.org/10.1088/0026-1394/62/1a/06004
- Primary Topic
- Advanced Radiotherapy Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00