Low background facility for ionizing radiation at LNE LNHB: Design and traceability

This paper presents the set up of a low background calibration facility aimed at calibrating environment survey meter. The facility is a lead box 100 cm x 100 cm x 125 cm with 2.5 cm thickness walls, the Caesium 137 source is located in a conical collimator including an internal cavity for the source. The ambient dose equivalent rate in the beam of this facility varies between 300 and 80 nSv/h. This quantity is traceable to the air kerma national references of LNE-LNHB. The standard uncertainty for the ambient dose equivalent rate is about 2% and the uncertainty of the calibration coefficient od a survey meter, depending meanly on the uncertainty if the background measure, is estimated between 5% and 10%.

Authors

Institutions

Publication Details

Journal
EPJ Web of Conferences
Published
2025-01-01
DOI
https://doi.org/10.1051/epjconf/202532311001
Primary Topic
Particle Detector Development and Performance
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Low background facility for ionizing radiation at LNE LNHB: Design and traceability

J.M. Bordy, O. Khattabi, C. Chardeur, K. Lahlal et al.
EPJ Web of Conferences
Particle Detector Development and Performance
article

Low background facility for ionizing radiation at LNE LNHB: Design and traceability

J.M. Bordy, O. Khattabi, C. Chardeur, K. Lahlal, J. Plangnard
article en

Abstract

This paper presents the set up of a low background calibration facility aimed at calibrating environment survey meter. The facility is a lead box 100 cm x 100 cm x 125 cm with 2.5 cm thickness walls, the Caesium 137 source is located in a conical collimator including an internal cavity for the source. The ambient dose equivalent rate in the beam of this facility varies between 300 and 80 nSv/h. This quantity is traceable to the air kerma national references of LNE-LNHB. The standard uncertainty for the ambient dose equivalent rate is about 2% and the uncertainty of the calibration coefficient od a survey meter, depending meanly on the uncertainty if the background measure, is estimated between 5% and 10%.

EPJ Web of ConferencesVol. 323
Commissariat à l'Énergie Atomique et aux Énergies Alternatives (FR), Université Paris-Saclay (FR), Laboratoire d'Intégration des Systèmes et des Technologies (FR), Laboratoire National Henri Becquerel (FR)
Openalex Percentile: Top 96%
Particle Detector Development and Performance
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.