Clinical implementation of patient-specific quality assurance for synthetic computed tomography
The present study proposes a robust PSQA method for MR-only planning. Method (A) is a valuable tool for identifying potential large outliers for Dmean deviations (> 5 %) and it is proposed as the routine PSQA. Method (B) can be used for pelvis cases to improve detection to the 2 % level if method (A) fails. If both (A) and (B) fail, method (C) can be used as a fall-back.