Can lead content be measured by handheld X‐ray fluorescence spectroscopy to meet <scp>REACH</scp> regulation?

Abstract Lead compounds have long been the most commonly used stabilizers for PVC products, but they are classified as toxic to reproduction category 1A, and their use is restricted in Europe. To promote recycling, the European Union adopted Regulation (EU) 2023/923, which permits the use of recycled rigid PVC (PVC‐U) containing up to 1.5% w/w of lead in specific products. This study compares the results obtained from two different handheld X‐Ray Fluorescence (HHXRF) devices on samples with known composition, also considering the matrix effect. This kind of evaluation is still missing in the literature. The measurements from both HHXRF devices show that the lead content falls within the 13% of relative inaccuracy range specified for lead in IEC 62321‐3‐1:2013, the standard currently used for determining lead in electrotechnical products. To assess whether the matrix effect can influence the measurements, two samples were produced by altering only the content of titanium dioxide. The results from both instruments were not significantly different, considering the 13% inaccuracy. In conclusion, HHXRF could be an effective method to determine the lead concentration in its usual dosage in the PVC‐U matrix. In case of inconclusive results based on this range, it is recommended to analyze the sample with another, more accurate technique, such as ICP. Highlights Statistical evaluation of accuracy of lead in samples with known composition. Evaluation of significant excess legal concentration. Prove infringements with fast and massive checks.

Authors

Institutions

Publication Details

Journal
Journal of Vinyl and Additive Technology
Published
2025-04-04
DOI
https://doi.org/10.1002/vnl.22220
Primary Topic
Heavy Metal Exposure and Toxicity
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Can lead content be measured by handheld X‐ray fluorescence spectroscopy to meet REACH regulation?

Costa Michele, Stefano Gardi, Polci Maria Letizia, Bonvicini Isotta et al.
Journal of Vinyl and Additive Technology
Heavy Metal Exposure and Toxicity
article

Can lead content be measured by handheld X‐ray fluorescence spectroscopy to meet REACH regulation?

Costa Michele, Stefano Gardi, Polci Maria Letizia, Bonvicini Isotta, Orsini Francesca, Sarti Gianluca
article en

Abstract

Abstract Lead compounds have long been the most commonly used stabilizers for PVC products, but they are classified as toxic to reproduction category 1A, and their use is restricted in Europe. To promote recycling, the European Union adopted Regulation (EU) 2023/923, which permits the use of recycled rigid PVC (PVC‐U) containing up to 1.5% w/w of lead in specific products. This study compares the results obtained from two different handheld X‐Ray Fluorescence (HHXRF) devices on samples with known composition, also considering the matrix effect. This kind of evaluation is still missing in the literature. The measurements from both HHXRF devices show that the lead content falls within the 13% of relative inaccuracy range specified for lead in IEC 62321‐3‐1:2013, the standard currently used for determining lead in electrotechnical products. To assess whether the matrix effect can influence the measurements, two samples were produced by altering only the content of titanium dioxide. The results from both instruments were not significantly different, considering the 13% inaccuracy. In conclusion, HHXRF could be an effective method to determine the lead concentration in its usual dosage in the PVC‐U matrix. In case of inconclusive results based on this range, it is recommended to analyze the sample with another, more accurate technique, such as ICP. Highlights Statistical evaluation of accuracy of lead in samples with known composition. Evaluation of significant excess legal concentration. Prove infringements with fast and massive checks.

Journal of Vinyl and Additive TechnologyVol. 31(4)
Istituto Superiore di Sanità (IT), University of Bologna (IT)
Openalex Percentile: Top 92%
Heavy Metal Exposure and Toxicity
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.