Mid-Infrared Spectrometer for Black Plastics Sorting Using a Broadband Uncooled Micro-Bolometer Array
We report the design, implementation and test of a Mid-Infrared spectrometer proof-of-concept that utilizes an uncooled micro-bolometer array, sensitive in the 3–14 µm spectral range, integrated in a conventional optical dispersive spectrometry setup. Such a spectrometer enables instantaneous measurements across this broad spectral range, comparable to that of a FTIR but with a more compact design and without moving parts. This makes it ideal for integration into portable, battery-powered devices such as handheld scanners. The Mid-IR range offers significant advantages over NIR-SWIR spectrometers, especially for organic compound analysis. A notable application for this instrument: plastic waste sorting—including black plastics—was tested with significant accuracy and effectiveness of plastic classification (on PP, PET and PE samples) with a very simple machine learning algorithm.
Authors
- Gabriel Jobert (ORCID: https://orcid.org/0009-0003-1246-4785)
- Xavier Brenière
Publication Details
- Journal
- Spectroscopy Journal
- Published
- 2025-04-03
- DOI
- https://doi.org/10.3390/spectroscj3020013
- Citations
- 1
- Primary Topic
- Industrial Vision Systems and Defect Detection
- Type
- article
- Field-Weighted Citation Impact
- 1.24
Funders
- Key Digital Technologies Joint Undertaking
- European Commission
- HORIZON EUROPE Framework Programme