A fully integrated whole-head helium OPM MEG: a performance assessment compared to cryogenic MEG

Magnetoencephalography (MEG) is a neuroimaging technique that measures neuronal activity at a millisecond scale. A few years ago, a new generation of MEG sensors emerged: optically pumped magnetometers (OPMs). The most common OPMs use alkali atoms as the sensing element. These alkali OPM sensors must be heated to approximately 150°C, in contrast to classical MEG sensors [superconducting quantum interference device MEG], which need to be cooled down to -269°C. This article focuses on a new kind of OPM that uses Helium-4 gas as the sensing element, which solves some disadvantages of alkali OPMs. <sup>4</sup>He-OPM sensors operate at room temperature, with negligible heat dissipation (10 mW) and thus do not need thermal insulation. They also offer a large dynamic range (±200 nT) and frequency bandwidth (2,000 Hz). The main goal of this study is to characterize the performance of a whole-head MEG system based on <sup>4</sup>He OPM sensors (<sup>4</sup>He OPM MEG). We first simulated different sensor configurations with three different numbers of channels and three different head sizes, from child to adult, in order to assess the signal-to-noise ratio and the source reconstruction accuracy. Experimental testing was also performed using a phantom to simulate brain magnetic activity. The simulation and experiments show equivalent detection capability and localization accuracy on both MEG systems. These results illustrate the benefit of <sup>4</sup>He OPM sensors that operate at room temperature and are positioned closer to the scalp.

Authors

Institutions

Publication Details

Journal
Frontiers in Medical Technology
Published
2025-04-04
DOI
https://doi.org/10.3389/fmedt.2025.1548260
Citations
10
Primary Topic
Atomic and Subatomic Physics Research
Type
article
Field-Weighted Citation Impact
12.47
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A fully integrated whole-head helium OPM MEG: a performance assessment compared to cryogenic MEG

Sébastien Daligault, Tjerk P. Gutteling, Denis Schwartz, Mathieu Bonnet et al.
10 citations
Frontiers in Medical Technology
Atomic and Subatomic Physics Research
12.47
article

A fully integrated whole-head helium OPM MEG: a performance assessment compared to cryogenic MEG

Sébastien Daligault, Tjerk P. Gutteling, Denis Schwartz, Mathieu Bonnet, Étienne Labyt
article en
10 citations

Abstract

Magnetoencephalography (MEG) is a neuroimaging technique that measures neuronal activity at a millisecond scale. A few years ago, a new generation of MEG sensors emerged: optically pumped magnetometers (OPMs). The most common OPMs use alkali atoms as the sensing element. These alkali OPM sensors must be heated to approximately 150°C, in contrast to classical MEG sensors [superconducting quantum interference device MEG], which need to be cooled down to -269°C. This article focuses on a new kind of OPM that uses Helium-4 gas as the sensing element, which solves some disadvantages of alkali OPMs. <sup>4</sup>He-OPM sensors operate at room temperature, with negligible heat dissipation (10 mW) and thus do not need thermal insulation. They also offer a large dynamic range (±200 nT) and frequency bandwidth (2,000 Hz). The main goal of this study is to characterize the performance of a whole-head MEG system based on <sup>4</sup>He OPM sensors (<sup>4</sup>He OPM MEG). We first simulated different sensor configurations with three different numbers of channels and three different head sizes, from child to adult, in order to assess the signal-to-noise ratio and the source reconstruction accuracy. Experimental testing was also performed using a phantom to simulate brain magnetic activity. The simulation and experiments show equivalent detection capability and localization accuracy on both MEG systems. These results illustrate the benefit of <sup>4</sup>He OPM sensors that operate at room temperature and are positioned closer to the scalp.

Frontiers in Medical TechnologyVol. 7
Université Claude Bernard Lyon 1 (FR), Centre National de la Recherche Scientifique (FR), Inserm (FR), Centre de Recherche en Neurosciences de Lyon (FR), Centre d'Exploration et de Recherche Médicale par Emission de Positons (FR)
Affordable and clean energy
Openalex Percentile: Top 1%
Atomic and Subatomic Physics Research
12.47
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.