Automated polarization rotation for multi-axis rotational-anisotropy second harmonic generation experiments

Rotational anisotropy second harmonic generation (RA-SHG) is a nonlinear optical technique used to probe the symmetry of condensed matter systems. Measuring the dependence of the SHG susceptibility on one or more external parameters, notably strain, field, temperature, or time delay, is an extremely powerful way to probe complex phases of quantum materials. Experimentally, extracting maximal information about the SHG susceptibility tensor requires measurements of S and P polarized input and output combinations, which naturally involves the rotation of the polarizers during data collection. For multi-axis experiments, this has proved challenging since polarization rotation is typically done manually. Automating this process eliminates labor constraints, reduces uncertainty due to low-frequency noise, and expands the type of multi-axis datasets that can be collected; however, it is difficult due to geometrical constraints within the setup. In this work, we design and implement low-cost, high-fidelity automated polarization rotators for use in multi-axis RA-SHG. These polarization rotators utilize an electrical slip ring to transfer power to the rotating RA-SHG optical setup as well as a miniature stepper motor to perform the polarization rotation. We demonstrate this automated system in time-resolved RA-SHG measurements in the non-centrosymmetric semiconductor GaAs. For the multi-axis measurements described above, this automated system permits data averaging over longer periods, vastly expedites data collection, and expands the setup measurement capability. This ultimately opens new frontiers in probing quantum materials using multiple tunable external parameters.

Authors

Institutions

Publication Details

Journal
Review of Scientific Instruments
Published
2025-04-01
DOI
https://doi.org/10.1063/5.0233827
Citations
1
Primary Topic
Spectroscopy and Quantum Chemical Studies
Type
article
Field-Weighted Citation Impact
1.25

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Automated polarization rotation for multi-axis rotational-anisotropy second harmonic generation experiments

Karna A. Morey, Nuh Gedik, Zongqi Shen, Bryan T. Fichera et al.
1 citations
Review of Scientific Instruments
Spectroscopy and Quantum Chemical Studies
1.25
article

Automated polarization rotation for multi-axis rotational-anisotropy second harmonic generation experiments

Karna A. Morey, Nuh Gedik, Zongqi Shen, Bryan T. Fichera, Baiqing Lv
article en
1 citations

Abstract

Rotational anisotropy second harmonic generation (RA-SHG) is a nonlinear optical technique used to probe the symmetry of condensed matter systems. Measuring the dependence of the SHG susceptibility on one or more external parameters, notably strain, field, temperature, or time delay, is an extremely powerful way to probe complex phases of quantum materials. Experimentally, extracting maximal information about the SHG susceptibility tensor requires measurements of S and P polarized input and output combinations, which naturally involves the rotation of the polarizers during data collection. For multi-axis experiments, this has proved challenging since polarization rotation is typically done manually. Automating this process eliminates labor constraints, reduces uncertainty due to low-frequency noise, and expands the type of multi-axis datasets that can be collected; however, it is difficult due to geometrical constraints within the setup. In this work, we design and implement low-cost, high-fidelity automated polarization rotators for use in multi-axis RA-SHG. These polarization rotators utilize an electrical slip ring to transfer power to the rotating RA-SHG optical setup as well as a miniature stepper motor to perform the polarization rotation. We demonstrate this automated system in time-resolved RA-SHG measurements in the non-centrosymmetric semiconductor GaAs. For the multi-axis measurements described above, this automated system permits data averaging over longer periods, vastly expedites data collection, and expands the setup measurement capability. This ultimately opens new frontiers in probing quantum materials using multiple tunable external parameters.

Review of Scientific InstrumentsVol. 96(4)
Shanghai Jiao Tong University (CN), Massachusetts Institute of Technology (US), Stanford University (US)
U.S. Department of Energy, Gordon and Betty Moore Foundation
Decent work and economic growth
Openalex Percentile: Top 23%
Spectroscopy and Quantum Chemical Studies
1.25
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.