Approval of retinal gene therapies in the US and Europe based on visual acuity and microperimetry
Microperimetry provides a sensitive and repeatable measure of retinal function, with mean sensitivity across the macula or central 16-point region offering a more reliable metric than single-point requirements recommended by the FDA. A 2.5dB gain in mean sensitivity is equivalent to a 13-letter increase in low-luminance ETDRS representing a clinically significant change and aligning closely with regulatory standards.
Authors
- Amandeep S. Josan (ORCID: https://orcid.org/0000-0003-1808-6546)
- Robert E. MacLaren (ORCID: https://orcid.org/0000-0002-3096-4682)
- Shabnam Raji (ORCID: https://orcid.org/0009-0003-6779-5194)
Institutions
- University of Oxford (GB)
- Oxford University Hospitals NHS Trust (GB)
Publication Details
- Journal
- Expert Review of Ophthalmology
- Published
- 2025-04-03
- DOI
- https://doi.org/10.1080/17469899.2025.2487544
- Primary Topic
- Retinal Development and Disorders
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- National Institute for Health and Care Research