In vitro and in vivo vein assessment of a novel vein visualizing device to improve first-time peripheral venous access

Inserting needles into veins is fundamental to medical care with up to 90% of inpatients requiring a peripheral intravenous catheter/cannula (PIVC) during their stay. Yet 40%–50% of PIVC insertions fail on the first attempt. Here, we present an easy-to-use novel vein visualizing ultrasound prototype device and data from in vitro and in vivo performance. Locational accuracy was determined through phantom simulated forearm veins, across variations of vein diameter (3–5 mm), depth (10–20 mm), and velocity (10–100 mm/s). Usability studies were conducted on nine clinicians to establish effectiveness and ease of use of the proposed prototype assisted cannulation workflow. Sensitivity of the prototype was demonstrated by scanning 80 forearm veins across 40 healthy volunteers. Our prototype's locational accuracy in simulated forearm veins is 0.21 mm ± 1.71 mm (s.d.) (97.7% agreement to the ground truth, p < .001). Usability studies found that 100% of users were able to handle the prototype in a sterile manner with minimal assistance. The sensitivity was excellent at finding veins (94%). In comparison, sensitivity of vein finding using landmark technique with torniquet (visible 46% and palpable 74%) were far inferior. Initial performance verification and validation studies presented suggest that the proposed ultrasound visualization method can simply and reliably help clinicians detect well-perfused veins at depth and visualize in the coronal view onboard the probe in alignment with the transducers. With improved ergonomics, the device has the potential to be an easy to use device for clinicians performing vascular access.

Authors

Institutions

Publication Details

Journal
WFUMB Ultrasound Open
Published
2025-04-04
DOI
https://doi.org/10.1016/j.wfumbo.2025.100085
Citations
1
Primary Topic
Central Venous Catheters and Hemodialysis
Type
article
Field-Weighted Citation Impact
2.52
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

In vitro and in vivo vein assessment of a novel vein visualizing device to improve first-time peripheral venous access

Nikhilesh Bappoo, Peter J. Carr, Barry J. Doyle, Matthew D. Redknap et al.
1 citations
WFUMB Ultrasound Open
Central Venous Catheters and Hemodialysis
2.52
article

In vitro and in vivo vein assessment of a novel vein visualizing device to improve first-time peripheral venous access

Nikhilesh Bappoo, Peter J. Carr, Barry J. Doyle, Matthew D. Redknap, Katherine Arenson, Michael D. Liddelow, Cara A. Boyce, Ellaby L. Hansen, N. M. Buckley, Phuoc Hao Ho
article en
1 citations

Abstract

Inserting needles into veins is fundamental to medical care with up to 90% of inpatients requiring a peripheral intravenous catheter/cannula (PIVC) during their stay. Yet 40%–50% of PIVC insertions fail on the first attempt. Here, we present an easy-to-use novel vein visualizing ultrasound prototype device and data from in vitro and in vivo performance. Locational accuracy was determined through phantom simulated forearm veins, across variations of vein diameter (3–5 mm), depth (10–20 mm), and velocity (10–100 mm/s). Usability studies were conducted on nine clinicians to establish effectiveness and ease of use of the proposed prototype assisted cannulation workflow. Sensitivity of the prototype was demonstrated by scanning 80 forearm veins across 40 healthy volunteers. Our prototype's locational accuracy in simulated forearm veins is 0.21 mm ± 1.71 mm (s.d.) (97.7% agreement to the ground truth, p < .001). Usability studies found that 100% of users were able to handle the prototype in a sterile manner with minimal assistance. The sensitivity was excellent at finding veins (94%). In comparison, sensitivity of vein finding using landmark technique with torniquet (visible 46% and palpable 74%) were far inferior. Initial performance verification and validation studies presented suggest that the proposed ultrasound visualization method can simply and reliably help clinicians detect well-perfused veins at depth and visualize in the coronal view onboard the probe in alignment with the transducers. With improved ergonomics, the device has the potential to be an easy to use device for clinicians performing vascular access.

WFUMB Ultrasound OpenVol. 3(1)
Griffith University (AU), The Kids Research Institute Australia (AU), The University of Western Australia (AU), Ollscoil na Gaillimhe – University of Galway (IE), Curtin University (AU), Harry Perkins Institute of Medical Research (AU), North Metropolitan Health Service (AU), Rio Tinto (Australia) (AU)
Openalex Percentile: Top 15%
Central Venous Catheters and Hemodialysis
2.52
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.