Impact of pullback speed on evaluation of lipid core plaque using near-infrared spectroscopy-intravascular ultrasound

Near-infrared spectroscopy-intravascular ultrasound (NIRS-IVUS) can assess the amount of lipid core plaque, and the latest generation catheter allows for imaging at several different pullback speeds. The purpose of this study was to investigate the impact of pullback speed on evaluation of lipid core plaque using the latest generation NIRS-IVUS catheter. Patients with coronary artery disease who underwent coronary angiography or percutaneous coronary intervention with NIRS-IVUS examination at 2 different pullback speeds (0.5 and 2.0 mm/sec) for the same vessel were enrolled. The values of lipid core burden index (LCBI) and maxLCBI4mm were compared between the images recorded with pullback speeds of 0.5 and 2.0 mm/sec, and match or mismatch of the location where maxLCBI4mm was detected was investigated. In a total of 286 NIRS-IVUS image sets, the location of maxLCBI4mm measured at 0.5 and 2.0 mm/sec was mismatched in 81 (28.3%) image sets. In comparison by vessel type, 20.8% were mismatched in LAD, 31.1% in LCx, and 37.8% in RCA (p = 0.023). In 205 image sets with matched maxLCBI4mm location, the maxLCBI4mm value at 2.0 mm/sec was significantly greater compared with that at 0.5 mm/sec (348.0 vs. 302.0 in median, p < 0.001). In a substantial proportion of patients, the location of maxLCBI4mm values measured at 0.5 and 2.0 mm/sec was mismatched. In the image sets with matched maxLCBI4mm location, the maxLCBI4mm value at 2.0 mm/sec was significantly greater compared with that at 0.5 mm/sec. These results should be noted when applying the cut-off values of maxLCBI4mm recorded at 2.0 mm/sec.

Authors

Institutions

Publication Details

Journal
Cardiovascular Intervention and Therapeutics
Published
2025-04-03
DOI
https://doi.org/10.1007/s12928-025-01124-7
Citations
2
Primary Topic
Coronary Interventions and Diagnostics
Type
article
Field-Weighted Citation Impact
4.01

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Impact of pullback speed on evaluation of lipid core plaque using near-infrared spectroscopy-intravascular ultrasound

Yuichi Saito, Takashi Iimori, Hideki Kitahara, Kazuya Tateishi et al.
2 citations
Cardiovascular Intervention and Therapeutics
Coronary Interventions and Diagnostics
4.01
article

Impact of pullback speed on evaluation of lipid core plaque using near-infrared spectroscopy-intravascular ultrasound

Yuichi Saito, Takashi Iimori, Hideki Kitahara, Kazuya Tateishi, Ken Kato, K. Takou, Tetsuharu Kasahara, Yoshio Kobayashi
article en
2 citations

Abstract

Near-infrared spectroscopy-intravascular ultrasound (NIRS-IVUS) can assess the amount of lipid core plaque, and the latest generation catheter allows for imaging at several different pullback speeds. The purpose of this study was to investigate the impact of pullback speed on evaluation of lipid core plaque using the latest generation NIRS-IVUS catheter. Patients with coronary artery disease who underwent coronary angiography or percutaneous coronary intervention with NIRS-IVUS examination at 2 different pullback speeds (0.5 and 2.0 mm/sec) for the same vessel were enrolled. The values of lipid core burden index (LCBI) and maxLCBI4mm were compared between the images recorded with pullback speeds of 0.5 and 2.0 mm/sec, and match or mismatch of the location where maxLCBI4mm was detected was investigated. In a total of 286 NIRS-IVUS image sets, the location of maxLCBI4mm measured at 0.5 and 2.0 mm/sec was mismatched in 81 (28.3%) image sets. In comparison by vessel type, 20.8% were mismatched in LAD, 31.1% in LCx, and 37.8% in RCA (p = 0.023). In 205 image sets with matched maxLCBI4mm location, the maxLCBI4mm value at 2.0 mm/sec was significantly greater compared with that at 0.5 mm/sec (348.0 vs. 302.0 in median, p < 0.001). In a substantial proportion of patients, the location of maxLCBI4mm values measured at 0.5 and 2.0 mm/sec was mismatched. In the image sets with matched maxLCBI4mm location, the maxLCBI4mm value at 2.0 mm/sec was significantly greater compared with that at 0.5 mm/sec. These results should be noted when applying the cut-off values of maxLCBI4mm recorded at 2.0 mm/sec.

Cardiovascular Intervention and TherapeuticsVol. 40(3)
Chiba University (JP), Chiba University Hospital (JP)
Chiba University
Good health and well-being
Openalex Percentile: Top 8%
Coronary Interventions and Diagnostics
4.01
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.