Advancing Myocardial T1 Mapping: A Comparative Study of the Frequency‐Independent MFA Sequence and Standard MOLLI

T1 mapping is essential for detecting myocardial changes, but standard methods like the MOLLI sequence are limited by heart rate dependency and sensitivity to motion artifacts. This study introduces the multiflip angle (MFA) sequence as a novel alternative, aiming to provide frequency-independent and robust T1 mapping, particularly in challenging cardiac conditions. The novel MFA sequence was validated using nickel (II) chloride phantoms and systematically compared with the standard MOLLI sequence in 20 healthy volunteers using a 1.5 Tesla Philips Achieva MRI system. T1 values were assessed at rest and under mild physical exertion to evaluate frequency dependency, measurement precision, and robustness to motion artifacts. The MFA sequence demonstrated robust frequency independence, with T1 values remaining stable across varying heart rates, unlike MOLLI, which exhibited a significant correlation between T1 values and heart rate (R = 0.52, p < 0.001), and sex (3% higher values in females; p = 0.044). Although both sequences showed no statistically significant age-related differences, MOLLI yielded more precise T1 measurements with lower variability compared to MFA. Additionally, MFA exhibited reduced susceptibility to motion artifacts, maintaining consistent values across myocardial regions and physiological conditions, particularly in basal segments where MOLLI showed greater variability. The MFA sequence offers a frequency-independent and motion-robust alternative to the MOLLI sequence for myocardial T1 mapping. Although the MOLLI sequence provides higher precision, MFA's stability across varying heart rates and resistance to motion artifacts positions it as a promising option, particularly for patients with arrhythmias or during stress testing. Further investigation is warranted to refine its clinical applications.

Authors

Institutions

Publication Details

Journal
NMR in Biomedicine
Published
2025-04-02
DOI
https://doi.org/10.1002/nbm.70031
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Advancing Myocardial T1 Mapping: A Comparative Study of the Frequency‐Independent MFA Sequence and Standard MOLLI

David M. Leistner, Vitali Koch, Mukaram Rana, Marco Ochs et al.
NMR in Biomedicine
Advanced MRI Techniques and Applications
article

Advancing Myocardial T1 Mapping: A Comparative Study of the Frequency‐Independent MFA Sequence and Standard MOLLI

David M. Leistner, Vitali Koch, Mukaram Rana, Marco Ochs, Simon S. Martin, Thomas J. Vogl, Sebastian M Haberkorn, Christina Seppi
article en

Abstract

T1 mapping is essential for detecting myocardial changes, but standard methods like the MOLLI sequence are limited by heart rate dependency and sensitivity to motion artifacts. This study introduces the multiflip angle (MFA) sequence as a novel alternative, aiming to provide frequency-independent and robust T1 mapping, particularly in challenging cardiac conditions. The novel MFA sequence was validated using nickel (II) chloride phantoms and systematically compared with the standard MOLLI sequence in 20 healthy volunteers using a 1.5 Tesla Philips Achieva MRI system. T1 values were assessed at rest and under mild physical exertion to evaluate frequency dependency, measurement precision, and robustness to motion artifacts. The MFA sequence demonstrated robust frequency independence, with T1 values remaining stable across varying heart rates, unlike MOLLI, which exhibited a significant correlation between T1 values and heart rate (R = 0.52, p < 0.001), and sex (3% higher values in females; p = 0.044). Although both sequences showed no statistically significant age-related differences, MOLLI yielded more precise T1 measurements with lower variability compared to MFA. Additionally, MFA exhibited reduced susceptibility to motion artifacts, maintaining consistent values across myocardial regions and physiological conditions, particularly in basal segments where MOLLI showed greater variability. The MFA sequence offers a frequency-independent and motion-robust alternative to the MOLLI sequence for myocardial T1 mapping. Although the MOLLI sequence provides higher precision, MFA's stability across varying heart rates and resistance to motion artifacts positions it as a promising option, particularly for patients with arrhythmias or during stress testing. Further investigation is warranted to refine its clinical applications.

NMR in BiomedicineVol. 38(5)
Goethe University Frankfurt (DE), German Centre for Cardiovascular Research (DE), University Hospital Frankfurt (DE), Sana Klinikum Offenbach (DE), Cardio-Pulmonary Institute (DE)
Good health and well-being
Openalex Percentile: Top 90%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.