Towards quantitative spectral mammography using triple-energy K-edge imaging and a two-pass reconstruction approach

This paper introduces a novel approach to quantitative contrast-enhanced spectral mammography (CESM) using triple energy K-edge imaging and a two-pass material decomposition method. The primary objective is to accurately determine local projected iodine concentration by leveraging the iodine K-edge in presence of glandular and adipose breast tissue. Addressing the ill-posed problem of established dual-energy methods, our approach accounts for three base materials with three non-redundant measurements. The proposed method involves two key steps. First, we estimate the local total thickness D via three material decomposition with the use of an iterative beam hardening correction, followed by strong denoising. Second, we use D as a constraint computing the recombined iodine image in classical dual-energy technique. In addition based on noise propagation the dose distribution and tube voltage for each image was optimized with respect to minimize standard deviation. Simulation studies using a numerical CIRS dual-energy phantom demonstrate thickness independent effective background cancellation, accurate reconstruction of relative projected iodine concentrations at moderate image noise levels. Sensitivity to remaining inaccuracies was simulated by adding scatter and deviation of used spectra for reconstruction and beam hardening correction. Furthermore, measurements were performed in laboratory system to test the feasibility. In conclusion, the proposed method provides a robust theoretical framework for accurately reconstructing projected iodine concentration images in CESM. For real measurements the method requires a highly accurate estimation of x-ray spectrum and knowledge of attenuation values.

Authors

Institutions

Publication Details

Published
2025-04-03
DOI
https://doi.org/10.1117/12.3048967
Citations
2
Primary Topic
Advanced X-ray Imaging Techniques
Type
article
Field-Weighted Citation Impact
3.77
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Towards quantitative spectral mammography using triple-energy K-edge imaging and a two-pass reconstruction approach

S. Kappler, Christopher Syben, Ferdinand Lück, Ludwig Ritschl et al.
2 citations
Advanced X-ray Imaging Techniques
3.77
article

Towards quantitative spectral mammography using triple-energy K-edge imaging and a two-pass reconstruction approach

S. Kappler, Christopher Syben, Ferdinand Lück, Ludwig Ritschl, Marcel Beister, Beatriz Pilar Garcia-Allende
article en
2 citations

Abstract

This paper introduces a novel approach to quantitative contrast-enhanced spectral mammography (CESM) using triple energy K-edge imaging and a two-pass material decomposition method. The primary objective is to accurately determine local projected iodine concentration by leveraging the iodine K-edge in presence of glandular and adipose breast tissue. Addressing the ill-posed problem of established dual-energy methods, our approach accounts for three base materials with three non-redundant measurements. The proposed method involves two key steps. First, we estimate the local total thickness D via three material decomposition with the use of an iterative beam hardening correction, followed by strong denoising. Second, we use D as a constraint computing the recombined iodine image in classical dual-energy technique. In addition based on noise propagation the dose distribution and tube voltage for each image was optimized with respect to minimize standard deviation. Simulation studies using a numerical CIRS dual-energy phantom demonstrate thickness independent effective background cancellation, accurate reconstruction of relative projected iodine concentrations at moderate image noise levels. Sensitivity to remaining inaccuracies was simulated by adding scatter and deviation of used spectra for reconstruction and beam hardening correction. Furthermore, measurements were performed in laboratory system to test the feasibility. In conclusion, the proposed method provides a robust theoretical framework for accurately reconstructing projected iodine concentration images in CESM. For real measurements the method requires a highly accurate estimation of x-ray spectrum and knowledge of attenuation values.

Siemens Healthcare (United States) (US), Siemens Healthineers (Germany) (DE)
Openalex Percentile: Top 8%
Advanced X-ray Imaging Techniques
3.77
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.