Novel Laser-Assisted Electrospray Ionization Mass Spectrometry (Laser-Assisted ESI-MS): A Sensitive Method for Determining Tetrabromobisphenol A and Its Derivative

Efficient ionization is essential for sensitive mass spectrometry (MS) analysis. Herein, a novel laser-assisted electrospray ionization (laser-assisted ESI) source was developed to efficiently ionize low-polar and thermal unstable compounds. By irradiating the electrospray nozzle with a simple and low-cost laser probe (450 nm, 500 mW), the sample droplets were stimulated with a laser beam as well as the high voltage of the electrospray, which significantly enhanced the ionization efficiency. In positive ionization mode, by further utilizing the reaction of Ag<sup>+</sup> and tetrabromobisphenol A bis(allyl ether) (TBBPA-BAE), the established laser-assisted ESI strategy was able to efficiently ionize the low-polar and thermal unstable TBBPA-BAE. Specifically, a limit of detection (LOD) of 14 ng L<sup>-1</sup>, linear range of 0.1-10 μg L<sup>-1</sup> (R<sup>2</sup> > 0.99), and relative standard deviations (RSDs) of 2.5% (n = 7, intraday) and 6.2% (n = 3 per day for 5 days, interday) were achieved. In negative ionization mode, laser-assisted ESI also improved the detection sensitivity of tetrabromobisphenol A (TBBPA), achieving a LOD of 3.3 ng L<sup>-1</sup>, linear range of 0.01-10 μg L<sup>-1</sup> (R<sup>2</sup> > 0.99), and RSDs of 4.7% (n = 7, intraday) and 7.3% (n = 3 per day for 5 days, interday). Compared to extractive electrospray ionization mass spectrometry (EESI-MS) and ESI-MS, the LODs achieved with laser-assisted ESI-MS were 54 and 16 times lower for the detection of TBBPA-BAE and TBBPA, respectively. Notably, deep purification and preconcentration were not required to accurately detect TBBPA and TBBPA-BAE in river water and wastewater treatment plant effluent. The spiked recoveries were between 88.0% and 106.0%, demonstrating the high reliability and practicality of this method.

Authors

Institutions

Publication Details

Journal
Analytical Chemistry
Published
2025-04-03
DOI
https://doi.org/10.1021/acs.analchem.5c00166
Citations
4
Primary Topic
Mass Spectrometry Techniques and Applications
Type
article
Field-Weighted Citation Impact
3.03

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Novel Laser-Assisted Electrospray Ionization Mass Spectrometry (Laser-Assisted ESI-MS): A Sensitive Method for Determining Tetrabromobisphenol A and Its Derivative

Ligang Hu, Xiaoxuan Han, Shuo Gao, Yong Tian et al.
4 citations
Analytical Chemistry
Mass Spectrometry Techniques and Applications
3.03
article

Novel Laser-Assisted Electrospray Ionization Mass Spectrometry (Laser-Assisted ESI-MS): A Sensitive Method for Determining Tetrabromobisphenol A and Its Derivative

Ligang Hu, Xiaoxuan Han, Shuo Gao, Yong Tian, Jianbo Shi, Fan Shi, Haonan Liu, Guibin Jiang
article en
4 citations

Abstract

Efficient ionization is essential for sensitive mass spectrometry (MS) analysis. Herein, a novel laser-assisted electrospray ionization (laser-assisted ESI) source was developed to efficiently ionize low-polar and thermal unstable compounds. By irradiating the electrospray nozzle with a simple and low-cost laser probe (450 nm, 500 mW), the sample droplets were stimulated with a laser beam as well as the high voltage of the electrospray, which significantly enhanced the ionization efficiency. In positive ionization mode, by further utilizing the reaction of Ag<sup>+</sup> and tetrabromobisphenol A bis(allyl ether) (TBBPA-BAE), the established laser-assisted ESI strategy was able to efficiently ionize the low-polar and thermal unstable TBBPA-BAE. Specifically, a limit of detection (LOD) of 14 ng L<sup>-1</sup>, linear range of 0.1-10 μg L<sup>-1</sup> (R<sup>2</sup> > 0.99), and relative standard deviations (RSDs) of 2.5% (n = 7, intraday) and 6.2% (n = 3 per day for 5 days, interday) were achieved. In negative ionization mode, laser-assisted ESI also improved the detection sensitivity of tetrabromobisphenol A (TBBPA), achieving a LOD of 3.3 ng L<sup>-1</sup>, linear range of 0.01-10 μg L<sup>-1</sup> (R<sup>2</sup> > 0.99), and RSDs of 4.7% (n = 7, intraday) and 7.3% (n = 3 per day for 5 days, interday). Compared to extractive electrospray ionization mass spectrometry (EESI-MS) and ESI-MS, the LODs achieved with laser-assisted ESI-MS were 54 and 16 times lower for the detection of TBBPA-BAE and TBBPA, respectively. Notably, deep purification and preconcentration were not required to accurately detect TBBPA and TBBPA-BAE in river water and wastewater treatment plant effluent. The spiked recoveries were between 88.0% and 106.0%, demonstrating the high reliability and practicality of this method.

Analytical ChemistryVol. 97(14)
Chinese Academy of Sciences (CN), Research Center for Eco-Environmental Sciences (CN), State Key Laboratory of Environmental Chemistry and Ecotoxicology, Qingdao University of Technology (CN)
National Natural Science Foundation of China, Natural Science Foundation of Shandong Province, Taishan Scholar Project of Shandong Province
Affordable and clean energy
Openalex Percentile: Top 10%
Mass Spectrometry Techniques and Applications
3.03
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.