Real-time blood gas management: evaluating quantum perfusion system’s accuracy against a standard blood gas analysis in CPB
The findings support the accuracy of the Quantum Perfusion System compared to the BGA standard, demonstrating the system's capability to provide accurate, continuous blood gas monitoring during CPB. However, further studies are necessary to strengthen and confirm these results across broader and more varied clinical scenarios, for these reason as recommended by the manufacturers, the quantum monitoring system should only be used as a trending device.