System accuracy evaluation of the new blood glucose monitoring meter “GLUCOCARD S onyx” beyond ISO 15197:2013/EN ISO 15197:2015 requirements and with new patient safety features

GLUCOCARD S onyx shows clinically satisfactory accuracy and reliability, even exceeding the ISO 15197:2013 criteria, for hypoglycemic cases with glucose critically low as <3.9 mmol/L (<70 mg/dL) and hyperglycemic cases with glucose ≥10.0 mmol/L (≥180 mg/dL). Healthcare organizations as well as manufacturers are aiming to offer new BGM systems that go beyond the ISO criteria and offer systems that can be consulted instead or besides CGM (Continuous Glucose Monitoring) in case of e.g. severe hypo- and/or hyperglycemic episodes. A MARD factor of 4.15% revealed an excellent system accuracy over the total clinically relevant glucose range. With additional user-friendly features, this BGM can be seen as a useful tool for efficient diabetes therapy, especially in the event of severe blood glucose fluctuations.

Authors

Institutions

Publication Details

Journal
Frontiers in Clinical Diabetes and Healthcare
Published
2025-04-03
DOI
https://doi.org/10.3389/fcdhc.2025.1465732
Primary Topic
Diabetes Management and Research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

System accuracy evaluation of the new blood glucose monitoring meter “GLUCOCARD S onyx” beyond ISO 15197:2013/EN ISO 15197:2015 requirements and with new patient safety features

Beate Saeger, Daisuke Azuma, Hisashi Okuda
Frontiers in Clinical Diabetes and Healthcare
Diabetes Management and Research
article

System accuracy evaluation of the new blood glucose monitoring meter “GLUCOCARD S onyx” beyond ISO 15197:2013/EN ISO 15197:2015 requirements and with new patient safety features

Beate Saeger, Daisuke Azuma, Hisashi Okuda
article en

Abstract

GLUCOCARD S onyx shows clinically satisfactory accuracy and reliability, even exceeding the ISO 15197:2013 criteria, for hypoglycemic cases with glucose critically low as <3.9 mmol/L (<70 mg/dL) and hyperglycemic cases with glucose ≥10.0 mmol/L (≥180 mg/dL). Healthcare organizations as well as manufacturers are aiming to offer new BGM systems that go beyond the ISO criteria and offer systems that can be consulted instead or besides CGM (Continuous Glucose Monitoring) in case of e.g. severe hypo- and/or hyperglycemic episodes. A MARD factor of 4.15% revealed an excellent system accuracy over the total clinically relevant glucose range. With additional user-friendly features, this BGM can be seen as a useful tool for efficient diabetes therapy, especially in the event of severe blood glucose fluctuations.

Frontiers in Clinical Diabetes and HealthcareVol. 6
Diabetes Australia (AU), Hewlett-Packard (Netherlands) (NL), Arkray (Japan) (JP)
Openalex Percentile: Top 91%
Diabetes Management and Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.