Noninvasive Glucose Measurements in Tissue Simulating Phantoms Using a Solid-State Near-Infrared Sensor

Benchmark data are reported for a solid-state laser-based near-infrared spectrometer designed for noninvasive measurements in human skin. These data were obtained using a set of aqueous phantoms composed of polystyrene beads, triton X-100, saline, and glucose. The performance of this prototype solid-state laser platform was compared to parallel results obtained with a Fourier-transform (FT) spectrometer. The fundamental spectroscopic performances of the two spectrometer systems were quantified by an analysis of 100% lines determined by ratioing back-to-back spectra collected over time for each phantom. Root mean square (RMS) noise levels were computed for each dataset and the median RMS noise levels were 327.8 µAU and 667.2 µAU for the FT spectrometer and prototype laser platform, respectively. The analytical utility of the solid-state laser platform was assessed through a series of leave-one-phantom-out partial least squares analyses. Results for the laser prototype data included a standard error of cross validation (SECV) of 7.82 mg/dL for an optimized PLS model with 10 factors over a spectral range of 1401-2238 nm. This compares favorably with the results from the FT spectrometer of an SECV of 6.62 mg/dL with 8 factors and a spectral range of 1551-2378 nm. The additional two PLS factors for the laser prototype were shown to be a consequence of its higher spectral noise. Selectivity of these PLS models was demonstrated by comparing models associated with correct and random glucose assignments to each spectrum. Overall, these findings benchmark the analytical utility of this solid-state laser prototype.

Authors

Institutions

Publication Details

Journal
Sensors
Published
2025-04-02
DOI
https://doi.org/10.3390/s25072238
Primary Topic
Spectroscopy Techniques in Biomedical and Chemical Research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Noninvasive Glucose Measurements in Tissue Simulating Phantoms Using a Solid-State Near-Infrared Sensor

Ariel Kauffman, Shuai Yu, Mark A. Arnold, Ruben Shakya
4 links
Sensors
Spectroscopy Techniques in Biomedical and Chemical Research
article

Noninvasive Glucose Measurements in Tissue Simulating Phantoms Using a Solid-State Near-Infrared Sensor

Ariel Kauffman, Shuai Yu, Mark A. Arnold, Ruben Shakya
article en

Abstract

Benchmark data are reported for a solid-state laser-based near-infrared spectrometer designed for noninvasive measurements in human skin. These data were obtained using a set of aqueous phantoms composed of polystyrene beads, triton X-100, saline, and glucose. The performance of this prototype solid-state laser platform was compared to parallel results obtained with a Fourier-transform (FT) spectrometer. The fundamental spectroscopic performances of the two spectrometer systems were quantified by an analysis of 100% lines determined by ratioing back-to-back spectra collected over time for each phantom. Root mean square (RMS) noise levels were computed for each dataset and the median RMS noise levels were 327.8 µAU and 667.2 µAU for the FT spectrometer and prototype laser platform, respectively. The analytical utility of the solid-state laser platform was assessed through a series of leave-one-phantom-out partial least squares analyses. Results for the laser prototype data included a standard error of cross validation (SECV) of 7.82 mg/dL for an optimized PLS model with 10 factors over a spectral range of 1401-2238 nm. This compares favorably with the results from the FT spectrometer of an SECV of 6.62 mg/dL with 8 factors and a spectral range of 1551-2378 nm. The additional two PLS factors for the laser prototype were shown to be a consequence of its higher spectral noise. Selectivity of these PLS models was demonstrated by comparing models associated with correct and random glucose assignments to each spectrum. Overall, these findings benchmark the analytical utility of this solid-state laser prototype.

SensorsVol. 25(7)
University of Iowa (US), Rockley Photonics (United States) (US)
Openalex Percentile: Top 94%
Spectroscopy Techniques in Biomedical and Chemical Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.