Randomized Cross-Over Study Comparing Three Models of Auto-Titrating Positive Airway Pressure Therapy for Obstructive Sleep Apnea

Objective: To compare pressure, average usage time, residual apnea-hypopnea index (AHI), and patient satisfaction across three models of auto-titrating positive airway pressure (APAP) (A, B, and C refers to APAP devices — APEX, Philips, and Hoffrichter, respectively). Materials and Methods: Seventy-five adult patients with an AHI of ≥15 events/h who were willing to use APAP were included and randomly assigned to three groups with different APAP sequences generated by computer randomization. After using each model at home for a week, patients returned to the clinic for data collection and switched to the next model. They completed questionnaires regarding their symptoms before and after the therapeutic session. Results: Data from 62 patients (43 males, 19 females) who completed the research protocol were analyzed. The average 90th–95th percentile pressures for APAP models A, B and C were 7.2 ± 2.2, 9.5 ± 2.6 and 7.4 ± 1.9 cmH2O, respectively (p < 0.001), with an intra-class correlation coefficient (ICC) of 0.52 (95% CI 0.22-0.70). In addition, average usage time and residual AHI differed significantly, with the highest values in model C. However, no significant differences were found in mean pressure or patient satisfaction across the three APAP models. Conclusion: The 90th–95th percentile pressures, average time usage, and residual AHI varied among APAP models, showing only moderate consistency. These findings suggest that careful consideration is required when selecting an APAP model for home use, as it may affect pressure determination and treatment outcomes.

Authors

Institutions

Publication Details

Journal
Siriraj Medical Journal/San Sirirat
Published
2025-04-01
DOI
https://doi.org/10.33192/smj.v77i4.272529
Citations
1
Primary Topic
Obstructive Sleep Apnea Research
Type
article
Field-Weighted Citation Impact
1.16

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Randomized Cross-Over Study Comparing Three Models of Auto-Titrating Positive Airway Pressure Therapy for Obstructive Sleep Apnea

Sarin Rungmanee, Wish Banhiran, Phawin Keskool, Wattanachai Chotinaiwattarakul et al.
1 citations
Siriraj Medical Journal/San Sirirat
Obstructive Sleep Apnea Research
1.16
article

Randomized Cross-Over Study Comparing Three Models of Auto-Titrating Positive Airway Pressure Therapy for Obstructive Sleep Apnea

Sarin Rungmanee, Wish Banhiran, Phawin Keskool, Wattanachai Chotinaiwattarakul, Sirithorn Kormongkolkul
article en
1 citations

Abstract

Objective: To compare pressure, average usage time, residual apnea-hypopnea index (AHI), and patient satisfaction across three models of auto-titrating positive airway pressure (APAP) (A, B, and C refers to APAP devices — APEX, Philips, and Hoffrichter, respectively). Materials and Methods: Seventy-five adult patients with an AHI of ≥15 events/h who were willing to use APAP were included and randomly assigned to three groups with different APAP sequences generated by computer randomization. After using each model at home for a week, patients returned to the clinic for data collection and switched to the next model. They completed questionnaires regarding their symptoms before and after the therapeutic session. Results: Data from 62 patients (43 males, 19 females) who completed the research protocol were analyzed. The average 90th–95th percentile pressures for APAP models A, B and C were 7.2 ± 2.2, 9.5 ± 2.6 and 7.4 ± 1.9 cmH2O, respectively (p < 0.001), with an intra-class correlation coefficient (ICC) of 0.52 (95% CI 0.22-0.70). In addition, average usage time and residual AHI differed significantly, with the highest values in model C. However, no significant differences were found in mean pressure or patient satisfaction across the three APAP models. Conclusion: The 90th–95th percentile pressures, average time usage, and residual AHI varied among APAP models, showing only moderate consistency. These findings suggest that careful consideration is required when selecting an APAP model for home use, as it may affect pressure determination and treatment outcomes.

Siriraj Medical Journal/San SiriratVol. 77(4)
Siriraj Hospital (TH), Mahidol University (TH)
Mahidol University, Faculty of Medicine Siriraj Hospital, Mahidol University
Openalex Percentile: Top 24%
Obstructive Sleep Apnea Research
1.16
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.