Evaluating the dosimetric and positioning accuracy of a deep learning based synthetic-CT model for liver radiotherapy treatment planning

<i>Background and purpose.</i>An MRI-only workflow requires synthetic computed tomography (sCT) images to enable dose calculation. This study evaluated the dosimetric and patient positioning accuracy of deep learning-generated sCT for liver radiotherapy.<i>Methods and materials.</i>sCT images were generated for eleven patients using a CycleGAN algorithm. Clinical volumetric modulated arc treatment plans (VMAT) were calculated on CT and recalculated on sCT, and dose differences were assessed using dose volume histogram (DVH). For position verification, the sCT images were validated as reference images to 4D cone beam computed tomography (4D CBCT) by calculating the translational and rotational differences between sCT and CT registrations to 4D CBCT.<i>Results.</i>The mean dose differences for the planning target volume (PTV) and organs at risk (OAR) between the CT and sCT plans were 0.0% and < 0.5%, respectively. For positioning verification, the systematic translational and rotational differences were < 0.5 mm and < 0.5°, respectively in all directions.<i>Conclusion.</i>This is the first study to validate a sCT model for liver cancer in terms of both dosimetry and patient positioning, marking a significant step in demonstrating the feasibility of an MRI-only workflow for treating liver cancer. The generated sCTs showed dosimetric differences within clinically acceptable levels and were successfully used as reference images for treatment position verification. This CycleGAN model is accessible through the research version of a commercial vendor, with potential for development as a clinical solution.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2025-04-02
DOI
https://doi.org/10.1088/2057-1976/adc818
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluating the dosimetric and positioning accuracy of a deep learning based synthetic-CT model for liver radiotherapy treatment planning

Bashar Al‐Qaisieh, Lamyaa Aljaafari, D. J. Bird, R. Speight et al.
Biomedical Physics & Engineering Express
Advanced Radiotherapy Techniques
article

Evaluating the dosimetric and positioning accuracy of a deep learning based synthetic-CT model for liver radiotherapy treatment planning

Bashar Al‐Qaisieh, Lamyaa Aljaafari, D. J. Bird, R. Speight, David L. Buckley, Sebastian Andersson
article en

Abstract

<i>Background and purpose.</i>An MRI-only workflow requires synthetic computed tomography (sCT) images to enable dose calculation. This study evaluated the dosimetric and patient positioning accuracy of deep learning-generated sCT for liver radiotherapy.<i>Methods and materials.</i>sCT images were generated for eleven patients using a CycleGAN algorithm. Clinical volumetric modulated arc treatment plans (VMAT) were calculated on CT and recalculated on sCT, and dose differences were assessed using dose volume histogram (DVH). For position verification, the sCT images were validated as reference images to 4D cone beam computed tomography (4D CBCT) by calculating the translational and rotational differences between sCT and CT registrations to 4D CBCT.<i>Results.</i>The mean dose differences for the planning target volume (PTV) and organs at risk (OAR) between the CT and sCT plans were 0.0% and < 0.5%, respectively. For positioning verification, the systematic translational and rotational differences were < 0.5 mm and < 0.5°, respectively in all directions.<i>Conclusion.</i>This is the first study to validate a sCT model for liver cancer in terms of both dosimetry and patient positioning, marking a significant step in demonstrating the feasibility of an MRI-only workflow for treating liver cancer. The generated sCTs showed dosimetric differences within clinically acceptable levels and were successfully used as reference images for treatment position verification. This CycleGAN model is accessible through the research version of a commercial vendor, with potential for development as a clinical solution.

Biomedical Physics & Engineering ExpressVol. 11(3)
University of Leeds (GB), King Saud bin Abdulaziz University for Health Sciences (SA), Leeds Teaching Hospitals NHS Trust (GB), RaySearch Laboratories (Sweden) (SE)
Good health and well-being
Openalex Percentile: Top 92%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.