Clinical Assessment of Potential Difference in Motion-Tracking Irradiation for Liver Tumors Using Radixact Synchrony®

This study aimed to advance the precision of motion-managed radiotherapy and provide robust information for clinical practice in terms of necessary margins using Radixact Synchrony® (Accuray, Sunnyvale, USA). A retrospective analysis was conducted on 88 irradiation sessions in nine cases of motion-tracking irradiation for liver tumors. Two gold markers were placed near the tumor, and a planning target volume (PTV) margin of 5 mm was set. The interruption threshold for potential difference (PD) (the statistical prediction value of the three-dimensional (3D) distance error from the predictive model) was set at 3.0 mm during treatment. PDvalues obtained before and during treatment, as well as the number of interruptions, were analyzed. A multiple regression analysis was performed to evaluate the influence of various planning parameters, which were obtained during four-dimensional (4D) treatment planning, on the 95th percentile value of PD (PD<sub>95</sub>). The mean PD<sub>95</sub> was 2.9 ± 0.6 mm. In the four cases where interruptions occurred, the average ± SD number of interruptions per fraction was 7.5 ± 4.6, 3.6 ± 2.5, 2.5 ± 1.5, and 1.0 ± 1.1, respectively. The coefficient of determination (adjusted R<sup>2</sup>) between PD<sub>95</sub> and planning parameters was 0.802 (p < 0.01), showing a positive correlation with respiratory motion amplitude (p = 0.017) and a negative correlation with inter-marker distance (p = 0.049). No treatment interruptions were experienced in 83.0% of a total of 88 fractions. To conclude, setting a PDinterruption threshold of 3.0 mm for motion-tracking irradiation using Radixact Synchrony® under metallic marker guidance for liver tumors was appropriate for minimizing treatment interruptions with sufficient dosimetric precision. In addition, inter-marker distance and respiratory motion amplitude were associated withPD<sub>95</sub>.

Authors

Institutions

Publication Details

Journal
Cureus
Published
2025-04-02
DOI
https://doi.org/10.7759/cureus.81598
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Clinical Assessment of Potential Difference in Motion-Tracking Irradiation for Liver Tumors Using Radixact Synchrony®

M. Tanooka, Shimpei Anami, Rina Muraoka, Hiroshi Doi et al.
Cureus
Advanced Radiotherapy Techniques
article

Clinical Assessment of Potential Difference in Motion-Tracking Irradiation for Liver Tumors Using Radixact Synchrony®

M. Tanooka, Shimpei Anami, Rina Muraoka, Hiroshi Doi, Takashi Shintani, Wataru Okada, Keisuke Sano
article en

Abstract

This study aimed to advance the precision of motion-managed radiotherapy and provide robust information for clinical practice in terms of necessary margins using Radixact Synchrony® (Accuray, Sunnyvale, USA). A retrospective analysis was conducted on 88 irradiation sessions in nine cases of motion-tracking irradiation for liver tumors. Two gold markers were placed near the tumor, and a planning target volume (PTV) margin of 5 mm was set. The interruption threshold for potential difference (PD) (the statistical prediction value of the three-dimensional (3D) distance error from the predictive model) was set at 3.0 mm during treatment. PDvalues obtained before and during treatment, as well as the number of interruptions, were analyzed. A multiple regression analysis was performed to evaluate the influence of various planning parameters, which were obtained during four-dimensional (4D) treatment planning, on the 95th percentile value of PD (PD<sub>95</sub>). The mean PD<sub>95</sub> was 2.9 ± 0.6 mm. In the four cases where interruptions occurred, the average ± SD number of interruptions per fraction was 7.5 ± 4.6, 3.6 ± 2.5, 2.5 ± 1.5, and 1.0 ± 1.1, respectively. The coefficient of determination (adjusted R<sup>2</sup>) between PD<sub>95</sub> and planning parameters was 0.802 (p < 0.01), showing a positive correlation with respiratory motion amplitude (p = 0.017) and a negative correlation with inter-marker distance (p = 0.049). No treatment interruptions were experienced in 83.0% of a total of 88 fractions. To conclude, setting a PDinterruption threshold of 3.0 mm for motion-tracking irradiation using Radixact Synchrony® under metallic marker guidance for liver tumors was appropriate for minimizing treatment interruptions with sufficient dosimetric precision. In addition, inter-marker distance and respiratory motion amplitude were associated withPD<sub>95</sub>.

CureusVol. 17(4)
Takarazuka City Hospital (JP), Kindai University (JP)
Openalex Percentile: Top 92%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.