Sonographic image texture features in muscle tissue-mimicking material reduce variability introduced by probe angle and gain settings compared to traditional echogenicity

Cost-effective and portable ultrasonography offers a promising approach for monitoring skeletal muscle damage and quality in many contexts. However, echogenicity analysis relies on precise transducer orientations and machine parameters, posing challenges for data pooling across different raters and settings. Muscle texture analysis offers a potential means of reducing inter-rater and machine-setting variability. Scans were assessed at nine angles, controlled using a custom transducer shell and software. Scans were performed three times, and different gains were applied. All scans were performed on a muscle tissue-mimicking phantom to eliminate biological variability. Intra-angle and intra-gain variability and internal consistency were assessed via coefficient of variation (CV%) and Cronbach's alpha (αc). Spearman's (ρ) correlations were employed to determine the relationship between echogenicity and each texture feature. Entropy (angle: CV=2.7-7.6%; gain: CV=10.5%; αc=0.86), and inverse difference moment (angle: CV=3.7-9.8%; gain: CV=16.5%; αc=0.87) were less variable than echogenicity (angle: CV=6.4-19.4%; gain: CV=39.0%; αc=0.82). Angular second moment (angle: CV=17.9-116.6%; gain: CV=71.6%; αc=0.68), contrast (angle: CV=7.8-14.7%; gain: CV=41.8%;αc=0.75), and correlation (angle: CV=9.0-13.5%; gain: CV=28.6%; αc=0.49) features were generally more variable. Entropy (ρ=0.82-0.98, p≤0.011) and inverse difference moment (ρ=-0.98--0.83, p≤0.008), were more strongly correlated with echogenicity than angular second moment (ρ=-0.98--0.77, p≤0.016), contrast (ρ=0.53-0.98, p≤0.15), and correlation (ρ=-0.25--0.19, p=0.520-0.631). Entropy and inverse difference moment features may allow data sharing between laboratory and clinical settings with ultrasound machine parameters and raters of varying skill levels. Clinical and mechanistic studies are required to determine if texture features can replace echogenicity assessments.

Authors

Institutions

Publication Details

Journal
European Journal of Translational Myology
Published
2025-04-01
DOI
https://doi.org/10.4081/ejtm.2025.13511
Citations
1
Primary Topic
Body Composition Measurement Techniques
Type
article
Field-Weighted Citation Impact
1.16
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Sonographic image texture features in muscle tissue-mimicking material reduce variability introduced by probe angle and gain settings compared to traditional echogenicity

Katie L. Boncella, Dustin J. Oranchuk, Michael O. Harris‐Love, Daniella Gonzalez-Rivera
1 citations
European Journal of Translational Myology
Body Composition Measurement Techniques
1.16
article

Sonographic image texture features in muscle tissue-mimicking material reduce variability introduced by probe angle and gain settings compared to traditional echogenicity

Katie L. Boncella, Dustin J. Oranchuk, Michael O. Harris‐Love, Daniella Gonzalez-Rivera
article en
1 citations

Abstract

Cost-effective and portable ultrasonography offers a promising approach for monitoring skeletal muscle damage and quality in many contexts. However, echogenicity analysis relies on precise transducer orientations and machine parameters, posing challenges for data pooling across different raters and settings. Muscle texture analysis offers a potential means of reducing inter-rater and machine-setting variability. Scans were assessed at nine angles, controlled using a custom transducer shell and software. Scans were performed three times, and different gains were applied. All scans were performed on a muscle tissue-mimicking phantom to eliminate biological variability. Intra-angle and intra-gain variability and internal consistency were assessed via coefficient of variation (CV%) and Cronbach's alpha (αc). Spearman's (ρ) correlations were employed to determine the relationship between echogenicity and each texture feature. Entropy (angle: CV=2.7-7.6%; gain: CV=10.5%; αc=0.86), and inverse difference moment (angle: CV=3.7-9.8%; gain: CV=16.5%; αc=0.87) were less variable than echogenicity (angle: CV=6.4-19.4%; gain: CV=39.0%; αc=0.82). Angular second moment (angle: CV=17.9-116.6%; gain: CV=71.6%; αc=0.68), contrast (angle: CV=7.8-14.7%; gain: CV=41.8%;αc=0.75), and correlation (angle: CV=9.0-13.5%; gain: CV=28.6%; αc=0.49) features were generally more variable. Entropy (ρ=0.82-0.98, p≤0.011) and inverse difference moment (ρ=-0.98--0.83, p≤0.008), were more strongly correlated with echogenicity than angular second moment (ρ=-0.98--0.77, p≤0.016), contrast (ρ=0.53-0.98, p≤0.15), and correlation (ρ=-0.25--0.19, p=0.520-0.631). Entropy and inverse difference moment features may allow data sharing between laboratory and clinical settings with ultrasound machine parameters and raters of varying skill levels. Clinical and mechanistic studies are required to determine if texture features can replace echogenicity assessments.

European Journal of Translational MyologyVol. 35(2)
Morpho (United States) (US), Geriatric Research Education and Clinical Center (US), VA Eastern Colorado Health Care System (US), University of Colorado Anschutz Medical Campus (US), University of Colorado Denver (US)
Openalex Percentile: Top 24%
Body Composition Measurement Techniques
1.16
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.