Patient-Specific QA Using Multidimensional Detectors in Single Isocenter Multitarget Stereotactic Radiotherapy for Multiple Brain Metastasis

Purpose: This study used ArcCHECK (AC) to investigate the connection between target- and treatment plan-dependent parameters and the gamma passing rate (GPR) to determine the characteristics of patient-specific quality assurance (PSQA) in single isocenter multitarget (SIMT) stereotactic radiotherapy (SRT). Methods: Twenty-four SIMT SRT treatment plans (92 targets) were evaluated. All treatment plans were developed using dynamic conformal arc therapy (DCAT), and the equivalent multi-leaf collimator (MLC) field size, distance from the isocenter (IC), and monitor/Gy were determined. The absolute dose error from the treatment planning system (TPS) was calculated using 92 targets and an ionization chamber detector. Gamma analysis was performed with AC, a multidimensional detector, and SNC patient software. Threshold values of 10% and 20%, absolute and relative dosimetry, global mode, and dose difference (DD)/distance-to-agreement (DTA) criteria of 3%/2 mm, 2%/2 mm, 2%/1 mm, and 1%/1 mm were employed. Differences in GPR were assessed for each condition. The correlation of GPR with treatment planning parameters and ionization chamber detector errors was investigated. Results: With stricter DD (2%/2 - 2%/1 mm), the GPR fell by an average of 18.6% (86.4% - 67.8%). The average equivalent MLC field size was moderately correlated with GPR and weakly correlated with the maximum error of the ionization chamber detector and MU/Gy, but not with the max target-IC distance. In the case of small targets, the dose in the center was relatively acceptable, but the lower dose range outside of the target was less responsive, frequently resulting in failures. Conclusions: In SIMT SRT, we investigated in detail the differences in GPR based on the AC’s gamma analysis conditions, as well as the relationship between the GPR and treatment planning parameters. The GPR was found to be significantly reduced when the average equivalent MLC field size was small, i.e., there were many small target sizes.

Authors

Publication Details

Journal
International Journal of Medical Physics Clinical Engineering and Radiation Oncology
Published
2025-01-01
DOI
https://doi.org/10.4236/ijmpcero.2025.142002
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Patient-Specific QA Using Multidimensional Detectors in Single Isocenter Multitarget Stereotactic Radiotherapy for Multiple Brain Metastasis

Hikaru Souda, Miyu Ishizawa, Hiraku Sato, Takeo Iwai et al.
1 links
International Journal of Medical Physics Clinical Engineering and Radiation Oncology
Advanced Radiotherapy Techniques
article

Patient-Specific QA Using Multidimensional Detectors in Single Isocenter Multitarget Stereotactic Radiotherapy for Multiple Brain Metastasis

Hikaru Souda, Miyu Ishizawa, Hiraku Sato, Takeo Iwai, Hongbo Chai, Mayumi Ichikawa, Yuya Miyasaka
article en

Abstract

Purpose: This study used ArcCHECK (AC) to investigate the connection between target- and treatment plan-dependent parameters and the gamma passing rate (GPR) to determine the characteristics of patient-specific quality assurance (PSQA) in single isocenter multitarget (SIMT) stereotactic radiotherapy (SRT). Methods: Twenty-four SIMT SRT treatment plans (92 targets) were evaluated. All treatment plans were developed using dynamic conformal arc therapy (DCAT), and the equivalent multi-leaf collimator (MLC) field size, distance from the isocenter (IC), and monitor/Gy were determined. The absolute dose error from the treatment planning system (TPS) was calculated using 92 targets and an ionization chamber detector. Gamma analysis was performed with AC, a multidimensional detector, and SNC patient software. Threshold values of 10% and 20%, absolute and relative dosimetry, global mode, and dose difference (DD)/distance-to-agreement (DTA) criteria of 3%/2 mm, 2%/2 mm, 2%/1 mm, and 1%/1 mm were employed. Differences in GPR were assessed for each condition. The correlation of GPR with treatment planning parameters and ionization chamber detector errors was investigated. Results: With stricter DD (2%/2 - 2%/1 mm), the GPR fell by an average of 18.6% (86.4% - 67.8%). The average equivalent MLC field size was moderately correlated with GPR and weakly correlated with the maximum error of the ionization chamber detector and MU/Gy, but not with the max target-IC distance. In the case of small targets, the dose in the center was relatively acceptable, but the lower dose range outside of the target was less responsive, frequently resulting in failures. Conclusions: In SIMT SRT, we investigated in detail the differences in GPR based on the AC’s gamma analysis conditions, as well as the relationship between the GPR and treatment planning parameters. The GPR was found to be significantly reduced when the average equivalent MLC field size was small, i.e., there were many small target sizes.

International Journal of Medical Physics Clinical Engineering and Radiation OncologyVol. 14(02)
Openalex Percentile: Top 92%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.