Uncertainty of Routine pH Measurements Evaluated by the Principle of Pooled Calibrations

In the digital age, taking logarithms of data is no longer necessary when assessing pH measurement uncertainties. The focus is now on using raw, unaltered data. Routine pH measurements often differ from those in reference labs, so quantifying these differences is necessary. The uncertainties of pH values at 25°C were evaluated using mass-by-mass calculations, along with comparisons between two pH meters. Four buffer systems and several beverage samples were tested. Expanded uncertainties (U) of up to U = 0.36 were observed for pH values between 2 and 11, significantly higher than those reported by manufacturers. While measured pH values were slightly higher than calculated ones, overlapping confidence intervals allowed the data to be combined. Due to significant uncertainties, the reliable pH range was limited to 1 < pH < 11, potentially narrower (1 < pH < 6.5). Routine pH measurement uncertainties did not match those based on the PoPC, with notably high CV values for proton activities at pH values below the buffers' pKa.

Authors

Institutions

Publication Details

Journal
IJCA (Indonesian Journal of Chemical Analysis)
Published
2025-03-30
DOI
https://doi.org/10.20885/ijca.vol8.iss1.art2
Primary Topic
Analytical Chemistry and Sensors
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Uncertainty of Routine pH Measurements Evaluated by the Principle of Pooled Calibrations

Mercy B. Menong, Jens Andersen
IJCA (Indonesian Journal of Chemical Analysis)
Analytical Chemistry and Sensors
article

Uncertainty of Routine pH Measurements Evaluated by the Principle of Pooled Calibrations

Mercy B. Menong, Jens Andersen
article en

Abstract

In the digital age, taking logarithms of data is no longer necessary when assessing pH measurement uncertainties. The focus is now on using raw, unaltered data. Routine pH measurements often differ from those in reference labs, so quantifying these differences is necessary. The uncertainties of pH values at 25°C were evaluated using mass-by-mass calculations, along with comparisons between two pH meters. Four buffer systems and several beverage samples were tested. Expanded uncertainties (U) of up to U = 0.36 were observed for pH values between 2 and 11, significantly higher than those reported by manufacturers. While measured pH values were slightly higher than calculated ones, overlapping confidence intervals allowed the data to be combined. Due to significant uncertainties, the reliable pH range was limited to 1 < pH < 11, potentially narrower (1 < pH < 6.5). Routine pH measurement uncertainties did not match those based on the PoPC, with notably high CV values for proton activities at pH values below the buffers' pKa.

IJCA (Indonesian Journal of Chemical Analysis)Vol. 8(1)
Botswana International University of Science and Technology (BW)
Openalex Percentile: Top 91%
Analytical Chemistry and Sensors
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.