Seismocardiograph Monitoring Using SMS Fiber Structure with PDMS Enclosure
This study presents a novel seismocardiography (SCG) monitoring system utilizing a singlemode-multimode-singlemode (SMS) fiber structure encased within a polydimethylsiloxane (PDMS) enclosure. The system employs a simple intensity-based interrogation method. Experimental characterization of the SMS fiber sensor demonstrated its sensitivity to vibrations, achieving an average calibration error of 1.63%. Validation against a gold-standard electrocardiogram (ECG) yielded an average error of 2.29%, representing a 1.27% accuracy improvement over existing fiber-optic SCG sensors (97.71% vs. 96.48%). This design offers a simplified and more robust alternative to current fiber-optic SCG sensors while maintaining high accuracy.
Authors
- Dhany Arifianto (ORCID: https://orcid.org/0000-0001-9805-598X)
- Frans Rizal Agustiyanto (ORCID: https://orcid.org/0000-0002-3103-708X)
- Selfi Stendafity
- Maulana Santoso
- Agus Muhamad Hatta (ORCID: https://orcid.org/0000-0001-5176-5933)
- Budi Susetyo Pikir (ORCID: https://orcid.org/0000-0003-0705-9462)
- Mahenda Radityo (ORCID: https://orcid.org/0009-0002-1689-5163)
Institutions
- Sepuluh Nopember Institute of Technology (ID)
- Universitas Airlangga (ID)
Publication Details
- Journal
- Journal of Biomedical Photonics & Engineering
- Published
- 2025-03-20
- DOI
- https://doi.org/10.18287/jbpe25.11.010303
- Primary Topic
- Non-Invasive Vital Sign Monitoring
- Type
- article
- Field-Weighted Citation Impact
- 0.00