A Method for Temporally Resolved Continuous Inline Measurement of Multiple Solute Concentrations With Microfluidic Spectroscopy

<i>Goal:</i> To develop a compact, real-time microfluidic spectroscopy system capable of simultaneously measuring the concentrations of multiple solutes flowing together through a single fluid pathway with high temporal resolution. <i>Methods:</i> The measurement system integrates a Z-flow cell and dual-wavelength LED light sources with a compact spectrophotometer. The experimental setup consisted of two clinical infusion pumps delivering distinct marker dyes through a common fluid pathway composed of a clinical manifold and a single lumen of a clinical intravascular catheter, while a third pump delivered an inert carrier fluid. Concentration measurements of the mixed dyes were performed at high-frequency sampling intervals, with dynamic pump rate adjustments to evaluate the system's ability to detect real-time changes in solute concentration. A MATLAB-based control application enabled automated data acquisition, processing, and system control to enhance experimental efficiency. <i>Results:</i> The system accurately measured solute concentrations, capturing temporal variations with high precision. It demonstrated high reproducibility with a standard error of the mean no larger than [Formula: see text] for Erioglaucine and [Formula: see text] for Tartrazine at steady state, and high accuracy with a maximum deviation of [Formula: see text] for Erioglaucine and [Formula: see text] for Tartrazine from the expected steady-state concentrations. <i>Conclusions:</i> This system enables continuous, real-time monitoring of multiple solutes in dynamic flow conditions, offering a portable solution with high sensitivity to temporal concentration changes-advancing beyond traditional static fluid measurement methods.

Authors

Institutions

Publication Details

Journal
IEEE Open Journal of Engineering in Medicine and Biology
Published
2025-01-01
DOI
https://doi.org/10.1109/ojemb.2025.3555807
Primary Topic
Water Quality Monitoring and Analysis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A Method for Temporally Resolved Continuous Inline Measurement of Multiple Solute Concentrations With Microfluidic Spectroscopy

Vineeth Chandran Suja, Robert A. Peterfreund, Nathaniel M. Sims, David Arney et al.
IEEE Open Journal of Engineering in Medicine and Biology
Water Quality Monitoring and Analysis
article

A Method for Temporally Resolved Continuous Inline Measurement of Multiple Solute Concentrations With Microfluidic Spectroscopy

Vineeth Chandran Suja, Robert A. Peterfreund, Nathaniel M. Sims, David Arney, Andrea Lorenzo Henri Sergio Detry
article en

Abstract

<i>Goal:</i> To develop a compact, real-time microfluidic spectroscopy system capable of simultaneously measuring the concentrations of multiple solutes flowing together through a single fluid pathway with high temporal resolution. <i>Methods:</i> The measurement system integrates a Z-flow cell and dual-wavelength LED light sources with a compact spectrophotometer. The experimental setup consisted of two clinical infusion pumps delivering distinct marker dyes through a common fluid pathway composed of a clinical manifold and a single lumen of a clinical intravascular catheter, while a third pump delivered an inert carrier fluid. Concentration measurements of the mixed dyes were performed at high-frequency sampling intervals, with dynamic pump rate adjustments to evaluate the system's ability to detect real-time changes in solute concentration. A MATLAB-based control application enabled automated data acquisition, processing, and system control to enhance experimental efficiency. <i>Results:</i> The system accurately measured solute concentrations, capturing temporal variations with high precision. It demonstrated high reproducibility with a standard error of the mean no larger than [Formula: see text] for Erioglaucine and [Formula: see text] for Tartrazine at steady state, and high accuracy with a maximum deviation of [Formula: see text] for Erioglaucine and [Formula: see text] for Tartrazine from the expected steady-state concentrations. <i>Conclusions:</i> This system enables continuous, real-time monitoring of multiple solutes in dynamic flow conditions, offering a portable solution with high sensitivity to temporal concentration changes-advancing beyond traditional static fluid measurement methods.

IEEE Open Journal of Engineering in Medicine and BiologyVol. 6
Harvard University (US), Harvard University Press (US), Massachusetts General Hospital (US), Federico II University Hospital (IT), University of Naples Federico II (IT)
Openalex Percentile: Top 92%
Water Quality Monitoring and Analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.