Simultaneous analysis of various anticancer drugs by supercritical fluid chromatography-mass spectrometry. Part I: Optimization of chromatographic separation

This work presents a generic SFC-MS method for the simultaneous analysis of 22 anticancer drugs (fluorouracil, busulfan, cyclophosphamide, cytarabine, dacarbazine, daunorubicin, docetaxel, doxorubicin, epirubicin, etoposide, gemcitabine, idarubicin, ifosfamide, irinotecan, methotrexate, paclitaxel, pemetrexed, raltitrexed, topotecan, treosulfan, vinblastine, vincristine). The separation conditions were optimized by screening nine stationary phases (2-picolylamine, bare hybrid silica, 2-ethylpyridine, fluoro-phenyl, octadecyl, diethylamine, diol, 1-aminoanthracene, zwitterionic modification), evaluating additives effects (2-5 % water, 20-50 mM ammonium formate, 0-1 mM ammonium fluoride), and adjusting the organic modifier composition (methanol, ethanol, isopropanol, acetonitrile). The optimized SFC-MS method successfully analyzed 22 anticancer drugs, along with 5 additional challenging compounds (azacitidine, mitomycin, cisplatin, oxaliplatin, carboplatin), in 12 min, using a diol column (100 × 3 mm, 1.7 µm) and a gradient of 2-100 % methanol containing 2 % water and 50 mM ammonium formate. To overcome overpressure generated by high organic solvent content, a backpressure gradient (110-150 bar) and a flow rate gradient (0.6-1.5 mL/min) were applied. The diol column was selected as the most promising based on five predefined chromatographic criteria. Additives with 5 % water or ammonium fluoride were excluded due to overpressure and signal loss, respectively. Increasing ammonium formate concentration improved peak symmetry by 29 %. For the organic modifier, pure methanol was chosen since ternary mixtures led to system overpressure without improving separation. Comparison with the LC-MS method using real samples confirmed the potential applicability of the SFC method, as the same trace compounds were detected with comparable concentrations. Sensitivity optimization and method validation will be discussed separately in a later paper.

Authors

Institutions

Publication Details

Journal
Journal of Pharmaceutical and Biomedical Analysis
Published
2025-03-29
DOI
https://doi.org/10.1016/j.jpba.2025.116838
Citations
2
Primary Topic
Safe Handling of Antineoplastic Drugs
Type
article
Field-Weighted Citation Impact
11.55
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Simultaneous analysis of various anticancer drugs by supercritical fluid chromatography-mass spectrometry. Part I: Optimization of chromatographic separation

Davy Guillarme, Pascal Bonnabry, Nathalie Nguyen, Serge Rudaz et al.
2 citations
Journal of Pharmaceutical and Biomedical Analysis
Safe Handling of Antineoplastic Drugs
11.55
article

Simultaneous analysis of various anticancer drugs by supercritical fluid chromatography-mass spectrometry. Part I: Optimization of chromatographic separation

Davy Guillarme, Pascal Bonnabry, Nathalie Nguyen, Serge Rudaz, Sandrine Fleury-Souverain
article en
2 citations

Abstract

This work presents a generic SFC-MS method for the simultaneous analysis of 22 anticancer drugs (fluorouracil, busulfan, cyclophosphamide, cytarabine, dacarbazine, daunorubicin, docetaxel, doxorubicin, epirubicin, etoposide, gemcitabine, idarubicin, ifosfamide, irinotecan, methotrexate, paclitaxel, pemetrexed, raltitrexed, topotecan, treosulfan, vinblastine, vincristine). The separation conditions were optimized by screening nine stationary phases (2-picolylamine, bare hybrid silica, 2-ethylpyridine, fluoro-phenyl, octadecyl, diethylamine, diol, 1-aminoanthracene, zwitterionic modification), evaluating additives effects (2-5 % water, 20-50 mM ammonium formate, 0-1 mM ammonium fluoride), and adjusting the organic modifier composition (methanol, ethanol, isopropanol, acetonitrile). The optimized SFC-MS method successfully analyzed 22 anticancer drugs, along with 5 additional challenging compounds (azacitidine, mitomycin, cisplatin, oxaliplatin, carboplatin), in 12 min, using a diol column (100 × 3 mm, 1.7 µm) and a gradient of 2-100 % methanol containing 2 % water and 50 mM ammonium formate. To overcome overpressure generated by high organic solvent content, a backpressure gradient (110-150 bar) and a flow rate gradient (0.6-1.5 mL/min) were applied. The diol column was selected as the most promising based on five predefined chromatographic criteria. Additives with 5 % water or ammonium fluoride were excluded due to overpressure and signal loss, respectively. Increasing ammonium formate concentration improved peak symmetry by 29 %. For the organic modifier, pure methanol was chosen since ternary mixtures led to system overpressure without improving separation. Comparison with the LC-MS method using real samples confirmed the potential applicability of the SFC method, as the same trace compounds were detected with comparable concentrations. Sensitivity optimization and method validation will be discussed separately in a later paper.

Journal of Pharmaceutical and Biomedical AnalysisVol. 261
University of Geneva (CH)
Good health and well-being
Openalex Percentile: Top 3%
Safe Handling of Antineoplastic Drugs
11.55
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.