Influence of variations in photon beam quality index on tissue inhomogeneity correction factors in radiation therapy treatment planning

This study was carried out to investigate how variations in the beam quality index (QI)-so-called tissue phantom ratio (TPR20,10) affect tissue inhomogeneity correction factors (ICFs) in external beam radiotherapy treatment plans. A total of 90 three-dimensional conformal radiotherapy (3DCRT) treatment plans for lung, gynecological, and prostate cancers and 15 stereotactic body radiation therapy (SBRT) plans for lung cancer were analyzed. For the 3DCRT plans, ICFs were evaluated across a range of beam QI values. For 6 MV photon beams, the TPR20,10 values were set at 0.670 ± 3%, while for 15 MV photon beams, the range was 0.760 ± 3%. SBRT plans were generated using two 6 MV photon beam configurations from a Varian TrueBeam accelerator – one employing a flattening filter (6 MV, TPR20,10 = 0.688) and the other operating in flattening filter-free (FFF) mode (6 MV FFF, TPR20,10 = 0.632). All dose calculations were performed using the Eclipse treatment planning system with the anisotropic analytical algorithm. For the calculations for 3DCRT plans, a 6.0% variation in the beam QI resulted in maximum differences in ICFs of 16.3% for lung cases with 6 MV beams and 12.5% for 15 MV beams. In gynecological and prostate cases, the ICF differences remained below 2.0% and 1.0%, respectively. In addition, for lung SBRT plans, a 5.6% discrepancy in TPR20,10 between 6 MV and 6 MV FFF beams led to ICF variations of <3.0%. These findings suggest that while variations in the beam QI significantly influence ICFs in lung cancer 3DCRT plans, their impact is less pronounced in gynecological and prostate treatments. This underscores the necessity of carefully accounting for beam quality variations during radiotherapy treatment planning to ensure accurate dose delivery and optimal patient outcomes.

Authors

Publication Details

Journal
Advances in Radiotherapy & Nuclear Medicine
Published
2025-03-27
DOI
https://doi.org/10.36922/arnm.6851
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Influence of variations in photon beam quality index on tissue inhomogeneity correction factors in radiation therapy treatment planning

Md. Akhtaruzzaman, Paweł Kukołowicz
Advances in Radiotherapy & Nuclear Medicine
Advanced Radiotherapy Techniques
article

Influence of variations in photon beam quality index on tissue inhomogeneity correction factors in radiation therapy treatment planning

Md. Akhtaruzzaman, Paweł Kukołowicz
article en

Abstract

This study was carried out to investigate how variations in the beam quality index (QI)-so-called tissue phantom ratio (TPR20,10) affect tissue inhomogeneity correction factors (ICFs) in external beam radiotherapy treatment plans. A total of 90 three-dimensional conformal radiotherapy (3DCRT) treatment plans for lung, gynecological, and prostate cancers and 15 stereotactic body radiation therapy (SBRT) plans for lung cancer were analyzed. For the 3DCRT plans, ICFs were evaluated across a range of beam QI values. For 6 MV photon beams, the TPR20,10 values were set at 0.670 ± 3%, while for 15 MV photon beams, the range was 0.760 ± 3%. SBRT plans were generated using two 6 MV photon beam configurations from a Varian TrueBeam accelerator – one employing a flattening filter (6 MV, TPR20,10 = 0.688) and the other operating in flattening filter-free (FFF) mode (6 MV FFF, TPR20,10 = 0.632). All dose calculations were performed using the Eclipse treatment planning system with the anisotropic analytical algorithm. For the calculations for 3DCRT plans, a 6.0% variation in the beam QI resulted in maximum differences in ICFs of 16.3% for lung cases with 6 MV beams and 12.5% for 15 MV beams. In gynecological and prostate cases, the ICF differences remained below 2.0% and 1.0%, respectively. In addition, for lung SBRT plans, a 5.6% discrepancy in TPR20,10 between 6 MV and 6 MV FFF beams led to ICF variations of <3.0%. These findings suggest that while variations in the beam QI significantly influence ICFs in lung cancer 3DCRT plans, their impact is less pronounced in gynecological and prostate treatments. This underscores the necessity of carefully accounting for beam quality variations during radiotherapy treatment planning to ensure accurate dose delivery and optimal patient outcomes.

Advances in Radiotherapy & Nuclear MedicineVol. 3(2)
Openalex Percentile: Top 92%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Influence of variations in photon beam quality index on tissue inhomogeneity correction factors in radiation therapy treatment planning — Md. Akhtaruzzaman, Paweł Kukołowicz · Advances in Radiotherapy & Nuclear Medicine (2025) | TGRS Research Map | TGRS