Facial Remote Photoplethysmography for Continuous Heart Rate, Stroke Volume, and Systemic Vascular Resistance Monitoring During Prolonged Warm/Cold Fluid Bolus Administration

Hemodynamic parameters—stroke volume (SV), systemic vascular resistance (SVR), and heart rate (HR)—are critical for clinical monitoring, particularly in emergency and critical care settings. This study evaluates the feasibility of a noninvasive, noncontact monitoring approach using imaging photoplethysmography (iPPG), which applies camera-based PPG and a signal processing pipeline implemented in MATLAB. A total of 25 prolonged video recordings (2500 min) were collected from 16 healthy volunteers at Odense University Hospital, while they received intravenous infusions of warm (<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$37~^{\circ }$ </tex-math></inline-formula>C) and cold (<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$15~^{\circ }$ </tex-math></inline-formula>C) Ringer’s lactate. To ensure data reliability, video quality and head motion were systematically analyzed. HR estimation using the plane-orthogonal-to-skin (POS) method achieved an average absolute error (avAE) of 4.28 bpm, with the best accuracy of 2.18 bpm, while the CHROM method yielded similar performance (4.27-bpm average error and 2.55-bpm best accuracy). SV and SVR demonstrated moderate correlation with reference measures (<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">${r} = {0.571}$ </tex-math></inline-formula> and <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">${r} = {0.596}$ </tex-math></inline-formula>, respectively) across five regions of interest. The best correlations for SV and SVR were <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">${r} = {0.846}$ </tex-math></inline-formula> and <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">${r} = {0.873}$ </tex-math></inline-formula>, respectively, indicating strong potential for accurate noncontact monitoring. These results suggest that iPPG can provide real-time cardiovascular insights during fluid therapy, with potential applications in telemedicine and mobile health. However, to robustly generalize the findings of this study, limitations such as interindividual variability and the need to include diverse patient populations should be considered. This study provides the first demonstration of iPPG feasibility for prolonged emergency fluid administration, paving the way for further research in dynamic clinical environments.

Authors

Institutions

Publication Details

Journal
IEEE Sensors Journal
Published
2025-03-28
DOI
https://doi.org/10.1109/jsen.2025.3553368
Citations
1
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
0.67

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Facial Remote Photoplethysmography for Continuous Heart Rate, Stroke Volume, and Systemic Vascular Resistance Monitoring During Prolonged Warm/Cold Fluid Bolus Administration

Peter Biesenbach, Mahdi Momeni, erik svendsen, Daniel Teichmann et al.
1 citations
IEEE Sensors Journal
Non-Invasive Vital Sign Monitoring
0.67
article

Facial Remote Photoplethysmography for Continuous Heart Rate, Stroke Volume, and Systemic Vascular Resistance Monitoring During Prolonged Warm/Cold Fluid Bolus Administration

Peter Biesenbach, Mahdi Momeni, erik svendsen, Daniel Teichmann, Mikkel Brabrand, Sophie Wuthe, Michaela Bitten Mølmer
article en
1 citations

Abstract

Hemodynamic parameters—stroke volume (SV), systemic vascular resistance (SVR), and heart rate (HR)—are critical for clinical monitoring, particularly in emergency and critical care settings. This study evaluates the feasibility of a noninvasive, noncontact monitoring approach using imaging photoplethysmography (iPPG), which applies camera-based PPG and a signal processing pipeline implemented in MATLAB. A total of 25 prolonged video recordings (2500 min) were collected from 16 healthy volunteers at Odense University Hospital, while they received intravenous infusions of warm (<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$37~^{\circ }$ </tex-math></inline-formula>C) and cold (<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$15~^{\circ }$ </tex-math></inline-formula>C) Ringer’s lactate. To ensure data reliability, video quality and head motion were systematically analyzed. HR estimation using the plane-orthogonal-to-skin (POS) method achieved an average absolute error (avAE) of 4.28 bpm, with the best accuracy of 2.18 bpm, while the CHROM method yielded similar performance (4.27-bpm average error and 2.55-bpm best accuracy). SV and SVR demonstrated moderate correlation with reference measures (<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">${r} = {0.571}$ </tex-math></inline-formula> and <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">${r} = {0.596}$ </tex-math></inline-formula>, respectively) across five regions of interest. The best correlations for SV and SVR were <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">${r} = {0.846}$ </tex-math></inline-formula> and <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">${r} = {0.873}$ </tex-math></inline-formula>, respectively, indicating strong potential for accurate noncontact monitoring. These results suggest that iPPG can provide real-time cardiovascular insights during fluid therapy, with potential applications in telemedicine and mobile health. However, to robustly generalize the findings of this study, limitations such as interindividual variability and the need to include diverse patient populations should be considered. This study provides the first demonstration of iPPG feasibility for prolonged emergency fluid administration, paving the way for further research in dynamic clinical environments.

IEEE Sensors JournalVol. 25(9)
University of Southern Denmark (DK), Maersk (Denmark) (DK), Odense University Hospital (DK), Region of Southern Denmark (DK), Hospital South West Jutland (DK)
Novo Nordisk Fonden
Good health and well-being
Openalex Percentile: Top 33%
Non-Invasive Vital Sign Monitoring
0.67
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.