Utility of a Card-Type Respiratory Rate Measuring Device for Spontaneously Breathing Patients

<b>Background/Objectives:</b> Accurate measurement of respiratory rate (RR) is critical for early detection of patient deterioration. A newly developed contactless card-type device measures RR and heart rate (HR) by detecting chest impedance changes. Although previously validated in mechanically ventilated patients with 15% RR accuracy, its performance in spontaneously breathing patients remains uncertain. <b>Methods:</b> This prospective observational study was conducted at the Department of Emergency and Critical Care Medicine, Shimane University Hospital, in December 2022. Patients admitted to the emergency center without invasive mechanical ventilation were enrolled. The card-type device was subsequently placed on the left chest, while the RR and HR were simultaneously recorded with an electrocardiogram monitor. Data from both devices were collected and compared. <b>Results:</b> Six patients were enrolled. The RR measurements from the card-type device were within a 10% difference from those measured using the standard monitor in four out of six cases. In two cases, the card-type device recorded an RR lower than that of the standard monitor, which coincided with periods of patient speaking. For HR, the card-type device was within a 10% margin of the standard monitor in two cases, but it underestimated HR in the remaining four cases, particularly during high tidal volumes or increased thoracic thickness. <b>Conclusions:</b> The contactless card-type device accurately measured respiratory rates within a 10% margin compared to standard monitors in most non-intubated patients, except during activities such as speaking. Further studies with larger sample sizes are warranted to confirm these findings and improve the device's performance.

Authors

Institutions

Publication Details

Journal
Diagnostics
Published
2025-03-28
DOI
https://doi.org/10.3390/diagnostics15070864
Primary Topic
Healthcare Technology and Patient Monitoring
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Utility of a Card-Type Respiratory Rate Measuring Device for Spontaneously Breathing Patients

Shun Takeda, Yoshiaki Iwashita, Satoru Nebuya, Shinya Sato et al.
Diagnostics
Healthcare Technology and Patient Monitoring
article

Utility of a Card-Type Respiratory Rate Measuring Device for Spontaneously Breathing Patients

Shun Takeda, Yoshiaki Iwashita, Satoru Nebuya, Shinya Sato, Satoshi Kawashima
article en

Abstract

<b>Background/Objectives:</b> Accurate measurement of respiratory rate (RR) is critical for early detection of patient deterioration. A newly developed contactless card-type device measures RR and heart rate (HR) by detecting chest impedance changes. Although previously validated in mechanically ventilated patients with 15% RR accuracy, its performance in spontaneously breathing patients remains uncertain. <b>Methods:</b> This prospective observational study was conducted at the Department of Emergency and Critical Care Medicine, Shimane University Hospital, in December 2022. Patients admitted to the emergency center without invasive mechanical ventilation were enrolled. The card-type device was subsequently placed on the left chest, while the RR and HR were simultaneously recorded with an electrocardiogram monitor. Data from both devices were collected and compared. <b>Results:</b> Six patients were enrolled. The RR measurements from the card-type device were within a 10% difference from those measured using the standard monitor in four out of six cases. In two cases, the card-type device recorded an RR lower than that of the standard monitor, which coincided with periods of patient speaking. For HR, the card-type device was within a 10% margin of the standard monitor in two cases, but it underestimated HR in the remaining four cases, particularly during high tidal volumes or increased thoracic thickness. <b>Conclusions:</b> The contactless card-type device accurately measured respiratory rates within a 10% margin compared to standard monitors in most non-intubated patients, except during activities such as speaking. Further studies with larger sample sizes are warranted to confirm these findings and improve the device's performance.

DiagnosticsVol. 15(7)
Shimane University (JP), Shimane University Hospital (JP)
Openalex Percentile: Top 90%
Healthcare Technology and Patient Monitoring
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Utility of a Card-Type Respiratory Rate Measuring Device for Spontaneously Breathing Patients — Shun Takeda, Yoshiaki Iwashita, et al. · Diagnostics (2025) | TGRS Research Map | TGRS