Comparison of The Optical Performance of Three Monofocal Plus Intraocular Lenses

Abstract Purpose To compare the real-life performance of three monofocal plus intraocular lenses (IOLs). Methods In this retrospective, single center, single surgeon case series, bilateral cataract patients with less than 1 diopter (D) of corneal astigmatism and no ocular comorbidities were included. Examinations were performed at 3 months postoperatively and included monocular and binocular uncorrected and corrected distance visual acuity, and distance-corrected (DC) visual acuity at 80, 60 and 40 cm, DC defocus curve with calculations of the area under the curve (AUC), contrast sensitivity and CATQUEST-9SF questionnaire. Results were all obtained with validated digital testing tablet applications. Results Among the 64 implanted eyes (32 patients) – 20 with Isopure123 (BVI Medical), 26 with Tecnis Eyhance (Johnson & Johnson Vision) and 18 with RayOne EMV (Rayner) IOLs – statistically significant differences (p ≤ 0.05) were found regarding 3 post-surgery CATQUEST-9SF categories and the defocus curve: between Eyhance and RayOne EMV (binocular) at −1.50 and −1.00D of defocus and in the near AUC segment, and between Isopure123 and RayOne EMV (monocular) for 0.00 and + 1.00D of defocus and in the far distance AUC segment. No statistically significant differences were found between the 3 IOLs for monocular and binocular uncorrected and corrected visual acuity at all distances tested. Conclusions All 3 IOL models provided comparable visual benefits and good performance at intermediate and near distance. A bigger sample size might detect more differences, as some trends were visible, such as differences in the defocus curve and patient-reported outcomes.

Authors

Institutions

Publication Details

Journal
Bratislavské lekárske listy/Bratislava medical journal
Published
2025-03-25
DOI
https://doi.org/10.1007/s44411-025-00108-6
Primary Topic
Ophthalmology and Visual Impairment Studies
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of The Optical Performance of Three Monofocal Plus Intraocular Lenses

Pavol Veselý, Alena Furdová, Jela Valášková, Natalia Parshina et al.
Bratislavské lekárske listy/Bratislava medical journal
Ophthalmology and Visual Impairment Studies
article

Comparison of The Optical Performance of Three Monofocal Plus Intraocular Lenses

Pavol Veselý, Alena Furdová, Jela Valášková, Natalia Parshina, Viktória Feketeová
article en

Abstract

Abstract Purpose To compare the real-life performance of three monofocal plus intraocular lenses (IOLs). Methods In this retrospective, single center, single surgeon case series, bilateral cataract patients with less than 1 diopter (D) of corneal astigmatism and no ocular comorbidities were included. Examinations were performed at 3 months postoperatively and included monocular and binocular uncorrected and corrected distance visual acuity, and distance-corrected (DC) visual acuity at 80, 60 and 40 cm, DC defocus curve with calculations of the area under the curve (AUC), contrast sensitivity and CATQUEST-9SF questionnaire. Results were all obtained with validated digital testing tablet applications. Results Among the 64 implanted eyes (32 patients) – 20 with Isopure123 (BVI Medical), 26 with Tecnis Eyhance (Johnson & Johnson Vision) and 18 with RayOne EMV (Rayner) IOLs – statistically significant differences (p ≤ 0.05) were found regarding 3 post-surgery CATQUEST-9SF categories and the defocus curve: between Eyhance and RayOne EMV (binocular) at −1.50 and −1.00D of defocus and in the near AUC segment, and between Isopure123 and RayOne EMV (monocular) for 0.00 and + 1.00D of defocus and in the far distance AUC segment. No statistically significant differences were found between the 3 IOLs for monocular and binocular uncorrected and corrected visual acuity at all distances tested. Conclusions All 3 IOL models provided comparable visual benefits and good performance at intermediate and near distance. A bigger sample size might detect more differences, as some trends were visible, such as differences in the defocus curve and patient-reported outcomes.

Bratislavské lekárske listy/Bratislava medical journalVol. 126(4)
University Hospital Bratislava (SK), Ocná klinika (SK), Comenius University Bratislava (SK)
Univerzita Komenského v Bratislave
Openalex Percentile: Top 91%
Ophthalmology and Visual Impairment Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.