The Influence of the Joint Volume on the Prevention of Impingement and Elbow-at-Side Rotations: Could the 36 mm Sphere with an Inferior Offset of 2 mm Be the New Gold Standard?

<b>Background</b>: Reverse shoulder arthroplasty (RSA) improves shoulder function in cases of glenohumeral osteoarthritis and rotator cuff arthropathy. The design of the glenosphere influences mobility and scapular impingement. This study evaluates the impact of joint volume on the range of motion (RoM) and identifies design modifications to enhance mobility while reducing the impingement risk. <b>Methods</b>: Thirty-four cadaveric shoulders were implanted with the Aequalis Reversed II<sup>®</sup> prosthesis in seven configurations: four with 36 mm spheres (centered, 2 mm eccentric, and lateralized by 5 mm and 7 mm) and three with 42 mm spheres (centered, and lateralized by 7 mm and 10 mm). The joint volumes (inferior, anteroinferior, and posteroinferior) were measured via 3D CT scans. The RoM in adduction and elbow-at-side rotations (IR1 and ER1) was recorded. A statistical analysis identified threshold joint volumes correlating with improved mobility. <b>Results</b>: Larger joint volumes correlated with enhanced mobility. The 42 mm spheres demonstrated better adduction and ER1 compared to those of the 36 mm spheres (<i>p</i> < 0.0001). An inferior volume > 5000 mm<sup>3</sup> and anteroinferior/posteroinferior volumes >2500 mm<sup>3</sup> were thresholds for significant mobility improvement. Lateralization (≥7 mm) or inferior eccentricity (2 mm) improved the mobility with the 36 mm spheres, with the 36 + 2 configuration offering a practical balance for smaller patients. <b>Conclusions</b>: Increased joint volume enhances mobility, particularly in adduction and elbow-at-side rotations. A sphere with a 2 mm inferior offset or a 42 sphere with 7 mm lateralization optimizes the RoM while minimizing impingement risks. Patient-specific considerations, including anatomy and soft tissue tension, remain essential for optimal prosthesis selection.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Medicine
Published
2025-03-28
DOI
https://doi.org/10.3390/jcm14072324
Citations
1
Primary Topic
Shoulder Injury and Treatment
Type
article
Field-Weighted Citation Impact
2.01
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The Influence of the Joint Volume on the Prevention of Impingement and Elbow-at-Side Rotations: Could the 36 mm Sphere with an Inferior Offset of 2 mm Be the New Gold Standard?

Marion Besnard, Ramy Samargandi, Julien Berhouet, Osamah Abualross
1 citations
Journal of Clinical Medicine
Shoulder Injury and Treatment
2.01
article

The Influence of the Joint Volume on the Prevention of Impingement and Elbow-at-Side Rotations: Could the 36 mm Sphere with an Inferior Offset of 2 mm Be the New Gold Standard?

Marion Besnard, Ramy Samargandi, Julien Berhouet, Osamah Abualross
article en
1 citations

Abstract

<b>Background</b>: Reverse shoulder arthroplasty (RSA) improves shoulder function in cases of glenohumeral osteoarthritis and rotator cuff arthropathy. The design of the glenosphere influences mobility and scapular impingement. This study evaluates the impact of joint volume on the range of motion (RoM) and identifies design modifications to enhance mobility while reducing the impingement risk. <b>Methods</b>: Thirty-four cadaveric shoulders were implanted with the Aequalis Reversed II<sup>®</sup> prosthesis in seven configurations: four with 36 mm spheres (centered, 2 mm eccentric, and lateralized by 5 mm and 7 mm) and three with 42 mm spheres (centered, and lateralized by 7 mm and 10 mm). The joint volumes (inferior, anteroinferior, and posteroinferior) were measured via 3D CT scans. The RoM in adduction and elbow-at-side rotations (IR1 and ER1) was recorded. A statistical analysis identified threshold joint volumes correlating with improved mobility. <b>Results</b>: Larger joint volumes correlated with enhanced mobility. The 42 mm spheres demonstrated better adduction and ER1 compared to those of the 36 mm spheres (<i>p</i> < 0.0001). An inferior volume > 5000 mm<sup>3</sup> and anteroinferior/posteroinferior volumes >2500 mm<sup>3</sup> were thresholds for significant mobility improvement. Lateralization (≥7 mm) or inferior eccentricity (2 mm) improved the mobility with the 36 mm spheres, with the 36 + 2 configuration offering a practical balance for smaller patients. <b>Conclusions</b>: Increased joint volume enhances mobility, particularly in adduction and elbow-at-side rotations. A sphere with a 2 mm inferior offset or a 42 sphere with 7 mm lateralization optimizes the RoM while minimizing impingement risks. Patient-specific considerations, including anatomy and soft tissue tension, remain essential for optimal prosthesis selection.

Journal of Clinical MedicineVol. 14(7)
Centre Hospitalier Universitaire de Tours (FR), University of Jeddah (SA)
Openalex Percentile: Top 17%
Shoulder Injury and Treatment
2.01
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.