Miniature Thermoelectric Cooler for Long-Term Stable Dehydration for On-Site Breath Analysis by Direct Inlet Photoionization Ion Mobility Spectrometry and Mass Spectrometry

Rapid and highly frequent measurements of exhaled breath hold significant value for disease diagnosis and exposure assessments. Direct inlet mass spectrometry and ion mobility spectrometry demonstrate great potential for on-site breath analysis, but their accuracy and sensitivity are notably affected by the high humidity in the breath samples. A miniature and continuous dehydration device based on dual-switching thermoelectric cold traps has been developed to efficiently reduce the humidity of breath gas from saturation to relative humidity of 4.3%, and a subcooling compensation method was adopted to eliminate the humidity fluctuations when switching the two cold traps to remove the condensed water in the trap, which could achieve a stable outlet humidity with RSD of 8.2% regardless of inlet humidity variations. The performance and profits of the miniature dehydration device for on-site analysis of breath samples have been tested by coupling with direct inlet photoionization ion mobility spectrometry and time-of-flight mass spectrometry. Trace breath acetone could be directly measured, and its concentration profiles over an 8 h period in fasting volunteers could be continuously monitored every 30 s. Furthermore, the <i>m</i>/<i>z</i> peaks from chlorinated hydrocarbons in exhaled breath samples were clearly identified after passing the dehydration device, which can be used for an occupational health assessment of the widely used halocarbon solvents.

Authors

Institutions

Publication Details

Journal
Analytical Chemistry
Published
2025-03-28
DOI
https://doi.org/10.1021/acs.analchem.5c00138
Citations
2
Primary Topic
Advanced Chemical Sensor Technologies
Type
article
Field-Weighted Citation Impact
1.35

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Miniature Thermoelectric Cooler for Long-Term Stable Dehydration for On-Site Breath Analysis by Direct Inlet Photoionization Ion Mobility Spectrometry and Mass Spectrometry

Yi Yu, Lei Hua, Huaiwen Cang, Yuanyuan Xie et al.
2 citations
Analytical Chemistry
Advanced Chemical Sensor Technologies
1.35
article

Miniature Thermoelectric Cooler for Long-Term Stable Dehydration for On-Site Breath Analysis by Direct Inlet Photoionization Ion Mobility Spectrometry and Mass Spectrometry

Yi Yu, Lei Hua, Huaiwen Cang, Yuanyuan Xie, Huimin Jiao, Jichun Jiang, Shan Deng, Haiyang Li, Zhihao Zhang, Hang Li, Qingyun Li
article en
2 citations

Abstract

Rapid and highly frequent measurements of exhaled breath hold significant value for disease diagnosis and exposure assessments. Direct inlet mass spectrometry and ion mobility spectrometry demonstrate great potential for on-site breath analysis, but their accuracy and sensitivity are notably affected by the high humidity in the breath samples. A miniature and continuous dehydration device based on dual-switching thermoelectric cold traps has been developed to efficiently reduce the humidity of breath gas from saturation to relative humidity of 4.3%, and a subcooling compensation method was adopted to eliminate the humidity fluctuations when switching the two cold traps to remove the condensed water in the trap, which could achieve a stable outlet humidity with RSD of 8.2% regardless of inlet humidity variations. The performance and profits of the miniature dehydration device for on-site analysis of breath samples have been tested by coupling with direct inlet photoionization ion mobility spectrometry and time-of-flight mass spectrometry. Trace breath acetone could be directly measured, and its concentration profiles over an 8 h period in fasting volunteers could be continuously monitored every 30 s. Furthermore, the <i>m</i>/<i>z</i> peaks from chlorinated hydrocarbons in exhaled breath samples were clearly identified after passing the dehydration device, which can be used for an occupational health assessment of the widely used halocarbon solvents.

Analytical ChemistryVol. 97(13)
Dalian Institute of Chemical Physics (CN), Dalian Medical University (CN), Chinese Academy of Sciences (CN), Second Affiliated Hospital of Dalian Medical University (CN), Shanghai Institute of Organic Chemistry (CN), University of Chinese Academy of Sciences (CN)
National Natural Science Foundation of China, Chinese Academy of Sciences, Natural Science Foundation of Liaoning Province, Dalian Medical University, National Key Research and Development Program of China
Openalex Percentile: Top 21%
Advanced Chemical Sensor Technologies
1.35
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.