Left Ventricular and Aortic Pressures Measured With Fluid‐Filled and Solid‐State Pressure Catheters: Similarities and Differences

Objective: To compare left ventricular (LV) and aortic (AO) pressures obtained using fluid‐filled and high‐fidelity solid‐state pressure catheters in subjects undergoing left heart catheterization. Materials and Methods: Twenty subjects scheduled for a left heart catheterization were enrolled and 18 subjects completed this IRB‐approved study. LV and AO pressures were obtained using fluid‐filled pressure catheter (standard‐of‐care) and high‐fidelity solid‐state pressure catheter synchronously. Pressure tracings were analyzed to measure LV systolic (LVSP), LV minimum‐diastolic (LVMDP), LV end‐diastolic (LVEDP), AO systolic (AOSP), and AO diastolic (AODP) pressures. Isovolumic contraction and relaxation rates (peak ± dp/dt) were derived from the pressure waveforms. Repeated measures of variance, post hoc tests with Bonferroni corrections, and Bland–Altman plots were used for comparisons. Results: A significant main effect of the pressure catheter was noted for LVSP, LVMDP, and AOSP ( p ≤ 0.025). The LVSP and AOSP measured with fluid‐filled pressure catheters were higher by 6.6 ± 6.9 mmHg and 4.6 ± 5.2 mmHg in comparison to solid‐state pressure catheter. In contrast, the LVMDP measurements were 3.5 ± 5.7 mmHg lower than the solid‐state pressure catheter measurements. The isovolumic contraction (66.4 ± 116.0 mmHg/s) and relaxation rates (60.5 ± 113.5 mmHg/s) were not significantly different between catheter systems after Bonferroni corrections for multiple comparisons ( p ≥ 0.06). The Bland–Altman analysis revealed a bias ranging from 0.8 to 6.6 mmHg. Conclusions: Differences in LVSP, LVMDP, and AOSP were noted between the catheter systems but not for other pressure values and contraction/relaxation rates. Fluid‐filled catheters overestimated true systolic pressures in the left ventricle and aorta. Trial Registration: ClinicalTrials.gov identifier: NCT03245255

Authors

Institutions

Publication Details

Journal
Journal of Interventional Cardiology
Published
2025-01-01
DOI
https://doi.org/10.1155/joic/9359365
Primary Topic
Cardiovascular Function and Risk Factors
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Left Ventricular and Aortic Pressures Measured With Fluid‐Filled and Solid‐State Pressure Catheters: Similarities and Differences

Marguerite Davis, David L. Fischman, Flemming Forsberg, Ira S. Cohen et al.
1 links
Journal of Interventional Cardiology
Cardiovascular Function and Risk Factors
article

Left Ventricular and Aortic Pressures Measured With Fluid‐Filled and Solid‐State Pressure Catheters: Similarities and Differences

Marguerite Davis, David L. Fischman, Flemming Forsberg, Ira S. Cohen, Priscilla Machado, Michael P. Savage, Cara Esposito, Praveen Mehrotra, Maureen E. McDonald, Jaydev K. Dave, Brian Osler
article en

Abstract

Objective: To compare left ventricular (LV) and aortic (AO) pressures obtained using fluid‐filled and high‐fidelity solid‐state pressure catheters in subjects undergoing left heart catheterization. Materials and Methods: Twenty subjects scheduled for a left heart catheterization were enrolled and 18 subjects completed this IRB‐approved study. LV and AO pressures were obtained using fluid‐filled pressure catheter (standard‐of‐care) and high‐fidelity solid‐state pressure catheter synchronously. Pressure tracings were analyzed to measure LV systolic (LVSP), LV minimum‐diastolic (LVMDP), LV end‐diastolic (LVEDP), AO systolic (AOSP), and AO diastolic (AODP) pressures. Isovolumic contraction and relaxation rates (peak ± dp/dt) were derived from the pressure waveforms. Repeated measures of variance, post hoc tests with Bonferroni corrections, and Bland–Altman plots were used for comparisons. Results: A significant main effect of the pressure catheter was noted for LVSP, LVMDP, and AOSP ( p ≤ 0.025). The LVSP and AOSP measured with fluid‐filled pressure catheters were higher by 6.6 ± 6.9 mmHg and 4.6 ± 5.2 mmHg in comparison to solid‐state pressure catheter. In contrast, the LVMDP measurements were 3.5 ± 5.7 mmHg lower than the solid‐state pressure catheter measurements. The isovolumic contraction (66.4 ± 116.0 mmHg/s) and relaxation rates (60.5 ± 113.5 mmHg/s) were not significantly different between catheter systems after Bonferroni corrections for multiple comparisons ( p ≥ 0.06). The Bland–Altman analysis revealed a bias ranging from 0.8 to 6.6 mmHg. Conclusions: Differences in LVSP, LVMDP, and AOSP were noted between the catheter systems but not for other pressure values and contraction/relaxation rates. Fluid‐filled catheters overestimated true systolic pressures in the left ventricle and aorta. Trial Registration: ClinicalTrials.gov identifier: NCT03245255

Journal of Interventional CardiologyVol. 2025(1)
Mayo Clinic (US), Thomas Jefferson University (US), Mayo Clinic in Arizona (US), Mayo Clinic in Florida (US)
American Heart Association, National Institutes of Health
Openalex Percentile: Top 91%
Cardiovascular Function and Risk Factors
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.