Assessment of the suitability of the dried blood spot (DBS) samples for detection of Treponemal antibodies using <i>in vitro</i> assays

Background & objectives This study was performed to assess the suitability of dried blood spot (DBS) samples for the serological screening of syphilis. Methods Two hundred paired DBS and plasma samples collected from six sexually transmitted infection (STI) clinics during the year 2023 were tested using three kits -Treponema pallidum haemagglutination assay (TPHA), Syphilis Total Ab and ErbaLisa Syphilis after standardization of the dilutions of the DBS elutes and considering the results of the paired plasma samples as a true status. Results The TPHA showed 89 per cent sensitivity and 100 per cent specificity, the EIA-Syphilis Total Ab. Kit showed 100 per cent sensitivity and 97 per cent specificity, whereas ErbaLisa syphilis showed 89 per cent sensitivity and 100 per cent specificity. However, one kit Syphilis ELISA (Oscar Medicare Pvt. Ltd., New Delhi) did not show agreement with the paired plasma samples at any dilution and was not considered suitable for the testing of DBS samples. The agreement between the plasma and DBS results was found to range from 94.5 to 98.5 per cent with a kappa agreement score of 0.89 for TPHA, ErbaLisa, and 0.97 for ELISA-Syphilis Total Ab. Interpretation & conclusions The findings of this study confirmed that the DBS samples can be used for the detection of anti-treponemal antibodies using the above-validated kits and thus may be a valuable tool in surveillance and epidemiological surveys conducted in India. The study also highlighted the need for validation of any plasma/serum syphilis antibody detection assay on DBS samples before using it on DBS samples.

Authors

Institutions

Publication Details

Journal
The Indian Journal of Medical Research
Published
2025-03-26
DOI
https://doi.org/10.25259/ijmr_1195_2024
Primary Topic
Syphilis Diagnosis and Treatment
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Assessment of the suitability of the dried blood spot (DBS) samples for detection of Treponemal antibodies using in vitro assays

Bhaswati Bandopadhyay, Pallavi Vidhate, Sachin Jadhav, Arati Mane et al.
1 links
The Indian Journal of Medical Research
Syphilis Diagnosis and Treatment
article

Assessment of the suitability of the dried blood spot (DBS) samples for detection of Treponemal antibodies using in vitro assays

Bhaswati Bandopadhyay, Pallavi Vidhate, Sachin Jadhav, Arati Mane, Shivali Kamal, Rohit Chawla, Sumathi Muralidhar, Suvarna Sane, Pramod Deoraj, Sagar E. Shirsath, Varsha Kale, Abdul Arif Khan, Hemlata Bhosale, Sangeeta Kulkarni, Smita Mishra, Madhuri Thakar, Pratik Kumar Jagtap, Sandhya Sawant, Bhawna Rao, Mayuri Borate
article en

Abstract

Background & objectives This study was performed to assess the suitability of dried blood spot (DBS) samples for the serological screening of syphilis. Methods Two hundred paired DBS and plasma samples collected from six sexually transmitted infection (STI) clinics during the year 2023 were tested using three kits -Treponema pallidum haemagglutination assay (TPHA), Syphilis Total Ab and ErbaLisa Syphilis after standardization of the dilutions of the DBS elutes and considering the results of the paired plasma samples as a true status. Results The TPHA showed 89 per cent sensitivity and 100 per cent specificity, the EIA-Syphilis Total Ab. Kit showed 100 per cent sensitivity and 97 per cent specificity, whereas ErbaLisa syphilis showed 89 per cent sensitivity and 100 per cent specificity. However, one kit Syphilis ELISA (Oscar Medicare Pvt. Ltd., New Delhi) did not show agreement with the paired plasma samples at any dilution and was not considered suitable for the testing of DBS samples. The agreement between the plasma and DBS results was found to range from 94.5 to 98.5 per cent with a kappa agreement score of 0.89 for TPHA, ErbaLisa, and 0.97 for ELISA-Syphilis Total Ab. Interpretation & conclusions The findings of this study confirmed that the DBS samples can be used for the detection of anti-treponemal antibodies using the above-validated kits and thus may be a valuable tool in surveillance and epidemiological surveys conducted in India. The study also highlighted the need for validation of any plasma/serum syphilis antibody detection assay on DBS samples before using it on DBS samples.

The Indian Journal of Medical ResearchVol. 161(2)
Safdarjang Hospital (IN), Topiwala National Medical College & BYL Nair Charitable Hospital (IN), Maulana Azad Medical College (IN), AIDS Society of India (IN), National AIDS Research Institute (IN), National AIDS Control Organisation (IN)
Openalex Percentile: Top 93%
Syphilis Diagnosis and Treatment
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.