New solar-powered device for the treatment of neonatal hyperbilirubinaemia in hardest-to-reach places tested in Nigeria

Background information The treatment of hyperbilirubinaemia in faraway remote locations of the world, where grid electricity and specialist doctors are unavailable, is hampered by a lack of appropriate technologies that could be operable by basic-trained healthcare workers. Hence the urgent need for the development of simplified low-cost medical devices that could be deployable to the hardest-to-reach remote locations of the world. Methods Low-cost stable elements, components and workpieces were targeted and applied in design to develop an applicable intensive phototherapy machine. Materials were sourced from in-country local markets to prototype the resultant device for capacity tests and technical and safety assessments. The device was crafted to be solely solar energy dependent for power sustainability and climate-change mitigation in remote locations prior to ethical approval and commencement of clinical trialling at University of Calabar Teaching Hospital Nigeria, to determine its efficiency in rapid breakdown of ‘severe’ serum bilirubin to a ‘mild’ benchmark level. Result The new device passed all technical safety assessments, delivering a total body irradiation treatment capacity of 45—158 µW/cm 2 /nm with 460 nm light wavelengths across various aspects of the body of a neonate. The device uses real-time solar power while self-banking its private reserve energy, which could last for over 27 hours when no real-time energy from the sun is available. All initial 11 patients enrolled during the first 3 months of trialling were safely and successfully treated within 17.8±8.7 irradiation-hours with stable vital signs and no injuries. Conclusion The new device is a potentially sustainable low-income and middle-income country solution for severe jaundice management.

Authors

Institutions

Publication Details

Journal
BMJ Innovations
Published
2025-01-01
DOI
https://doi.org/10.1136/bmjinnov-2024-001286
Citations
3
Primary Topic
Neonatal Health and Biochemistry
Type
article
Field-Weighted Citation Impact
10.48
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

New solar-powered device for the treatment of neonatal hyperbilirubinaemia in hardest-to-reach places tested in Nigeria

Martin Madu Meremikwu, J J Udo, Hippolite O. Amadi, Joanah Moses Ikobah et al.
3 citations
BMJ Innovations
Neonatal Health and Biochemistry
10.48
article

New solar-powered device for the treatment of neonatal hyperbilirubinaemia in hardest-to-reach places tested in Nigeria

Martin Madu Meremikwu, J J Udo, Hippolite O. Amadi, Joanah Moses Ikobah, Sunny Ochigbo, Abdulrasheed Jimoh, Obot E. Antia-Obong, E. B. Adams, Charity Ojong Ajake, Leo-Stanley Chinonso Muoneke, Esther U Asi
article en
3 citations

Abstract

Background information The treatment of hyperbilirubinaemia in faraway remote locations of the world, where grid electricity and specialist doctors are unavailable, is hampered by a lack of appropriate technologies that could be operable by basic-trained healthcare workers. Hence the urgent need for the development of simplified low-cost medical devices that could be deployable to the hardest-to-reach remote locations of the world. Methods Low-cost stable elements, components and workpieces were targeted and applied in design to develop an applicable intensive phototherapy machine. Materials were sourced from in-country local markets to prototype the resultant device for capacity tests and technical and safety assessments. The device was crafted to be solely solar energy dependent for power sustainability and climate-change mitigation in remote locations prior to ethical approval and commencement of clinical trialling at University of Calabar Teaching Hospital Nigeria, to determine its efficiency in rapid breakdown of ‘severe’ serum bilirubin to a ‘mild’ benchmark level. Result The new device passed all technical safety assessments, delivering a total body irradiation treatment capacity of 45—158 µW/cm 2 /nm with 460 nm light wavelengths across various aspects of the body of a neonate. The device uses real-time solar power while self-banking its private reserve energy, which could last for over 27 hours when no real-time energy from the sun is available. All initial 11 patients enrolled during the first 3 months of trialling were safely and successfully treated within 17.8±8.7 irradiation-hours with stable vital signs and no injuries. Conclusion The new device is a potentially sustainable low-income and middle-income country solution for severe jaundice management.

BMJ InnovationsVol. 11(1)
University of Calabar (NG), Imperial College London (GB)
Openalex Percentile: Top 3%
Neonatal Health and Biochemistry
10.48
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.