Accelerated diffusion-weighted imaging of the prostate employing echo planar imaging with compressed SENSE based reconstruction

Aim was to evaluate accelerated diffusion-weighted imaging (DWI) of the prostate using echo planar imaging with compressed SENSE based reconstruction (EPICS) and assess its performance in comparison to conventional DWI with parallel imaging. In this single-center, prospective study, 35 men with clinically suspected prostate cancer underwent prostate MRI at 3T. In each patient, two different DWI sequences, one with 3 b-values (b = 100, 400, 800s/mm²) for ADC-calculation and one with b = 1500s/mm², were acquired with conventional SENSE and with EPICS. Quantitative evaluation was done by regions-of-interest (ROIs) analysis of prostate lesions and normal appearing peripheral zones (PZ). Apparent contrast-to-noise (aCNR) and apparent signal-to-noise ratios (aSNR) were calculated. Mean ADC and coefficient of variation (CV) of ADC were compared. For qualitative assessment, artifacts, lesion conspicuity, and overall image quality were rated using a 5-point-Likert-scale (1: nondiagnostic to 5: excellent). Additionally, the Prostate Imaging Reporting and Data System (PIRADS 2.1) was rated for DWI. The average total scan time reduction with EPICS was 43%. Quantitative analysis showed no significant differences between conventional SENSE and EPICS, neither for aSNR<sub>Lesion</sub> (e.g. b1500<sub>conv</sub>: 24.37 ± 10.28 vs. b1500<sub>EPICS</sub>: 24.08 ± 12.2; p = 0.98) and aCNR<sub>Lesion</sub> (e.g. b1500<sub>conv</sub>:9.53 ± 7.22 vs. b1500<sub>EPICS</sub>:8.88 ± 6.16; p = 0.55) nor for aSNR<sub>PZ</sub> (e.g. b1500<sub>conv</sub>:15.18 ± 6.48 vs. b1500<sub>EPICS</sub>: 15 ± 7.4; p = 0.94). Rating of artifacts, lesion conspicuity, overall image quality and PIRADS-scores yielded comparable results for the two techniques (e.g. lesion conspicuity for ADC<sub>conv</sub>: 4(2-5) vs. ADC<sub>EPICS</sub> 4(2-5); p = 0.99 and for b1500<sub>conv</sub>: 4(2-5) vs. b1500<sub>EPICS</sub> 4(2-5); p = 0.25). Overall, accelerated DWI of the prostate using EPICS significantly reduced acquisition time without compromising image quality compared to conventional DWI.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2025-03-25
DOI
https://doi.org/10.1038/s41598-025-94777-6
Citations
1
Primary Topic
Prostate Cancer Diagnosis and Treatment
Type
article
Field-Weighted Citation Impact
1.52
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Accelerated diffusion-weighted imaging of the prostate employing echo planar imaging with compressed SENSE based reconstruction

Petra Mürtz, Julian Alexander Luetkens, Claus Christian Pieper, Alexander Isaak et al.
1 citations
Scientific Reports
Prostate Cancer Diagnosis and Treatment
1.52
article

Accelerated diffusion-weighted imaging of the prostate employing echo planar imaging with compressed SENSE based reconstruction

Petra Mürtz, Julian Alexander Luetkens, Claus Christian Pieper, Alexander Isaak, Kilian Weiss, Leon Bischoff, Christoph Katemann, Yannik C. Layer, Barbara Daria Wichtmann
article en
1 citations

Abstract

Aim was to evaluate accelerated diffusion-weighted imaging (DWI) of the prostate using echo planar imaging with compressed SENSE based reconstruction (EPICS) and assess its performance in comparison to conventional DWI with parallel imaging. In this single-center, prospective study, 35 men with clinically suspected prostate cancer underwent prostate MRI at 3T. In each patient, two different DWI sequences, one with 3 b-values (b = 100, 400, 800s/mm²) for ADC-calculation and one with b = 1500s/mm², were acquired with conventional SENSE and with EPICS. Quantitative evaluation was done by regions-of-interest (ROIs) analysis of prostate lesions and normal appearing peripheral zones (PZ). Apparent contrast-to-noise (aCNR) and apparent signal-to-noise ratios (aSNR) were calculated. Mean ADC and coefficient of variation (CV) of ADC were compared. For qualitative assessment, artifacts, lesion conspicuity, and overall image quality were rated using a 5-point-Likert-scale (1: nondiagnostic to 5: excellent). Additionally, the Prostate Imaging Reporting and Data System (PIRADS 2.1) was rated for DWI. The average total scan time reduction with EPICS was 43%. Quantitative analysis showed no significant differences between conventional SENSE and EPICS, neither for aSNR<sub>Lesion</sub> (e.g. b1500<sub>conv</sub>: 24.37 ± 10.28 vs. b1500<sub>EPICS</sub>: 24.08 ± 12.2; p = 0.98) and aCNR<sub>Lesion</sub> (e.g. b1500<sub>conv</sub>:9.53 ± 7.22 vs. b1500<sub>EPICS</sub>:8.88 ± 6.16; p = 0.55) nor for aSNR<sub>PZ</sub> (e.g. b1500<sub>conv</sub>:15.18 ± 6.48 vs. b1500<sub>EPICS</sub>: 15 ± 7.4; p = 0.94). Rating of artifacts, lesion conspicuity, overall image quality and PIRADS-scores yielded comparable results for the two techniques (e.g. lesion conspicuity for ADC<sub>conv</sub>: 4(2-5) vs. ADC<sub>EPICS</sub> 4(2-5); p = 0.99 and for b1500<sub>conv</sub>: 4(2-5) vs. b1500<sub>EPICS</sub> 4(2-5); p = 0.25). Overall, accelerated DWI of the prostate using EPICS significantly reduced acquisition time without compromising image quality compared to conventional DWI.

Scientific ReportsVol. 15(1)
University Hospital Bonn (DE), Philips (Germany) (DE)
Openalex Percentile: Top 20%
Prostate Cancer Diagnosis and Treatment
1.52
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.