Wireless Electromagnetic Tracking System for Wireless Capsule Endoscope

Electromagnetic tracking (EMT) is widely employed in image-guided surgical procedures, enabling precise tracking of the position and orientation of medical instruments. However, the traditional EMT system needs wires to be led out from the transmitting side to the receiving side, rendering it unsuitable for in vivo biomedical implants. This study proposes a wireless EMT system tailored for the wireless capsule endoscope (WCE) application, which includes: 1) elucidation of fundamental comparison between “wired” EMT and “wireless” EMT; 2) devising a corresponding wireless EMT algorithm; and 3) a small-sized, low-power hardware architecture for the receiving side of EMT. The receiving circuit is about the size of the capsule, and consumes 4.73 mW excluding the wireless communication module. Experiment results show the effectiveness of the proposed wireless EMT. Within a tracking region of <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$20\times 15\times 13$ </tex-math></inline-formula> cm3, the position error is <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$3.24~\pm ~0.88$ </tex-math></inline-formula> mm and the orientation error is <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$1.95~\pm ~0.79^{\circ }$ </tex-math></inline-formula>, with an update rate of 1 Hz.

Authors

Institutions

Publication Details

Journal
IEEE Transactions on Instrumentation and Measurement
Published
2025-01-01
DOI
https://doi.org/10.1109/tim.2025.3550218
Citations
1
Primary Topic
Gastrointestinal Bleeding Diagnosis and Treatment
Type
article
Field-Weighted Citation Impact
1.81

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Wireless Electromagnetic Tracking System for Wireless Capsule Endoscope

Yuming Fu, Yong‐Xin Guo, Xiaoyang Wu, Yi Li et al.
1 citations
IEEE Transactions on Instrumentation and Measurement
Gastrointestinal Bleeding Diagnosis and Treatment
1.81
article

Wireless Electromagnetic Tracking System for Wireless Capsule Endoscope

Yuming Fu, Yong‐Xin Guo, Xiaoyang Wu, Yi Li, Ming Liu
article en
1 citations

Abstract

Electromagnetic tracking (EMT) is widely employed in image-guided surgical procedures, enabling precise tracking of the position and orientation of medical instruments. However, the traditional EMT system needs wires to be led out from the transmitting side to the receiving side, rendering it unsuitable for in vivo biomedical implants. This study proposes a wireless EMT system tailored for the wireless capsule endoscope (WCE) application, which includes: 1) elucidation of fundamental comparison between “wired” EMT and “wireless” EMT; 2) devising a corresponding wireless EMT algorithm; and 3) a small-sized, low-power hardware architecture for the receiving side of EMT. The receiving circuit is about the size of the capsule, and consumes 4.73 mW excluding the wireless communication module. Experiment results show the effectiveness of the proposed wireless EMT. Within a tracking region of <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$20\times 15\times 13$ </tex-math></inline-formula> cm3, the position error is <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$3.24~\pm ~0.88$ </tex-math></inline-formula> mm and the orientation error is <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$1.95~\pm ~0.79^{\circ }$ </tex-math></inline-formula>, with an update rate of 1 Hz.

IEEE Transactions on Instrumentation and MeasurementVol. 74
National University of Singapore (SG), City University of Hong Kong (HK), Suzhou Research Institute (CN)
City University of Hong Kong
Affordable and clean energy
Openalex Percentile: Top 19%
Gastrointestinal Bleeding Diagnosis and Treatment
1.81
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.