A Technologist’s Vigilance: Identifying and Correcting a Cotton Ball-Induced MRI Artefact to Prevent Misdiagnosis in Pediatric Patients

Wrap-around artefacts in magnetic resonance imaging (MRI) are common, typically caused by anatomical structures outside the Field of View (FOV) overlapping into the imaging area. This paper reports a rare source of wrap-around artifact, a wet cotton ball, whose image was inadvertently included in cranial imaging, potentially leading to misdiagnosis as pathological conditions such as otomastoiditis, postoperative changes, or intracranial hemorrhage. Hence, there is a need to enhance MRI technologists' awareness of such artifacts. We analyzed four consecutive cases of cranial MRI scans with similar artifacts, all located on the right side of the brain but in different regions. On axial T2-weighted images (T2WI) and T2 fluid-attenuated inversion recovery (T2FLAIR) sequences, the artifacts appeared as oval-shaped, relatively well-defined heterogeneous high signals. On axial T1-weighted images (T1WI), signal intensity and border clarity varied, with artifact locations slightly more lateral compared to T2WI and T2FLAIR. The artifacts in the first three cases were not identified by the MRI technologists. However, in the fourth case, a patient with head trauma, the MRI technologist noticed inconsistencies in the artifact characteristics across different sequences, as well as similarities to a previous case of cranial trauma, which raised suspicion of an artifact. Systematic image analysis and equipment inspection subsequently revealed a wet cotton ball attached to the outer surface of the head-and-neck coil, inadvertently left there by a radiology nurse during preparation for a previous case. Upon removal of the cotton ball and rescanning the sequences with artifacts in cases 3 and 4, the artifacts disappeared, confirming the wet cotton ball as the source. Additionally, upon review, similar artifacts were found in the first two cases but overlooked due to various reasons. The aim of this study is to improve MRI technologists' recognition of artifacts caused by non-metallic foreign objects, avoiding misdiagnosis, and to prompt us to refine our examination protocols and enhance radiology nurses' awareness of MRI safety.

Authors

Institutions

Publication Details

Journal
Cureus
Published
2025-03-26
DOI
https://doi.org/10.7759/cureus.81252
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A Technologist’s Vigilance: Identifying and Correcting a Cotton Ball-Induced MRI Artefact to Prevent Misdiagnosis in Pediatric Patients

Nan Zhou, Longping Liu, Xiaoli Zheng, Weiguo Cao
Cureus
Advanced MRI Techniques and Applications
article

A Technologist’s Vigilance: Identifying and Correcting a Cotton Ball-Induced MRI Artefact to Prevent Misdiagnosis in Pediatric Patients

Nan Zhou, Longping Liu, Xiaoli Zheng, Weiguo Cao
article en

Abstract

Wrap-around artefacts in magnetic resonance imaging (MRI) are common, typically caused by anatomical structures outside the Field of View (FOV) overlapping into the imaging area. This paper reports a rare source of wrap-around artifact, a wet cotton ball, whose image was inadvertently included in cranial imaging, potentially leading to misdiagnosis as pathological conditions such as otomastoiditis, postoperative changes, or intracranial hemorrhage. Hence, there is a need to enhance MRI technologists' awareness of such artifacts. We analyzed four consecutive cases of cranial MRI scans with similar artifacts, all located on the right side of the brain but in different regions. On axial T2-weighted images (T2WI) and T2 fluid-attenuated inversion recovery (T2FLAIR) sequences, the artifacts appeared as oval-shaped, relatively well-defined heterogeneous high signals. On axial T1-weighted images (T1WI), signal intensity and border clarity varied, with artifact locations slightly more lateral compared to T2WI and T2FLAIR. The artifacts in the first three cases were not identified by the MRI technologists. However, in the fourth case, a patient with head trauma, the MRI technologist noticed inconsistencies in the artifact characteristics across different sequences, as well as similarities to a previous case of cranial trauma, which raised suspicion of an artifact. Systematic image analysis and equipment inspection subsequently revealed a wet cotton ball attached to the outer surface of the head-and-neck coil, inadvertently left there by a radiology nurse during preparation for a previous case. Upon removal of the cotton ball and rescanning the sequences with artifacts in cases 3 and 4, the artifacts disappeared, confirming the wet cotton ball as the source. Additionally, upon review, similar artifacts were found in the first two cases but overlooked due to various reasons. The aim of this study is to improve MRI technologists' recognition of artifacts caused by non-metallic foreign objects, avoiding misdiagnosis, and to prompt us to refine our examination protocols and enhance radiology nurses' awareness of MRI safety.

CureusVol. 17(3)
Shenzhen Children's Hospital (CN)
Openalex Percentile: Top 91%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.