A pre-clinical PET scanner based on a monolithic annulus of scintillator (AnnPET): construction and NU4-2008 performance testing

<i>Objective.</i>In the past several decades, numerous positron emission tomography (PET) scanners of various designs have been constructed for use in pre-clinical studies. Our group is investigating use of a monolithic annulus of scintillator, instead of the traditional arrays of discrete scintillator elements or individual detectors that utilize continuous blocks of scintillator, to construct a novel pre-clinical PET scanner.<i>Approach.</i>This scanner, called AnnPET, is based on a fourteen-faceted annulus of lutetium yttrium orthosilicate with an inner diameter of 6 cm and length of 7.2 cm. Each facet is populated with four specially constructed 4 × 4 arrays of 4 mm × 4 mm multi-pixel photon counters .To cool and temperature stabilize these devices, the scanner gantry is immersed in dielectric fluid. Positioning of events in the scintillator is accomplished with the application of deep-residual convolutional neural network. The scanner's performance was assessed using the NEMA NU4-2008 protocols.<i>Results.</i>Full-width-at-half-maximum (FWHM) of the images of a point source reconstructed with the single slice rebinned filtered backprojection (SSRB-FBP) algorithm at 5 mm from the center of the scanner are: 1.40 mm (radial), 1.38 mm (tangential) and 1.40 mm (axial). At 18 mm from scanner center (edge of the scanner's inner bore) the FWHMs are: 1.62 mm (radial), 1.43 mm (tangential) and 1.48 mm (axial) FWHM. Peak detection sensitivity is 9.5% (0.086 cps Bq<sup>-1</sup>). Peak noise equivalent count rate is 234 kcps at 14.4 MBq.<i>Significance.</i>Overall, testing of the AnnPET system demonstrated very promising performance results for a pre-clinical PET scanner based on a single, cooled annulus of monolithic scintillator used with neural networks. Continued development of the system is planned.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2025-03-25
DOI
https://doi.org/10.1088/1361-6560/adc537
Citations
1
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
1.36

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A pre-clinical PET scanner based on a monolithic annulus of scintillator (AnnPET): construction and NU4-2008 performance testing

Raymond R. Raylman, Gangadhar Jaliparthi, Peter Martone, Alexander Stolin
1 citations
Physics in Medicine and Biology
Medical Imaging Techniques and Applications
1.36
article

A pre-clinical PET scanner based on a monolithic annulus of scintillator (AnnPET): construction and NU4-2008 performance testing

Raymond R. Raylman, Gangadhar Jaliparthi, Peter Martone, Alexander Stolin
article en
1 citations

Abstract

<i>Objective.</i>In the past several decades, numerous positron emission tomography (PET) scanners of various designs have been constructed for use in pre-clinical studies. Our group is investigating use of a monolithic annulus of scintillator, instead of the traditional arrays of discrete scintillator elements or individual detectors that utilize continuous blocks of scintillator, to construct a novel pre-clinical PET scanner.<i>Approach.</i>This scanner, called AnnPET, is based on a fourteen-faceted annulus of lutetium yttrium orthosilicate with an inner diameter of 6 cm and length of 7.2 cm. Each facet is populated with four specially constructed 4 × 4 arrays of 4 mm × 4 mm multi-pixel photon counters .To cool and temperature stabilize these devices, the scanner gantry is immersed in dielectric fluid. Positioning of events in the scintillator is accomplished with the application of deep-residual convolutional neural network. The scanner's performance was assessed using the NEMA NU4-2008 protocols.<i>Results.</i>Full-width-at-half-maximum (FWHM) of the images of a point source reconstructed with the single slice rebinned filtered backprojection (SSRB-FBP) algorithm at 5 mm from the center of the scanner are: 1.40 mm (radial), 1.38 mm (tangential) and 1.40 mm (axial). At 18 mm from scanner center (edge of the scanner's inner bore) the FWHMs are: 1.62 mm (radial), 1.43 mm (tangential) and 1.48 mm (axial) FWHM. Peak detection sensitivity is 9.5% (0.086 cps Bq<sup>-1</sup>). Peak noise equivalent count rate is 234 kcps at 14.4 MBq.<i>Significance.</i>Overall, testing of the AnnPET system demonstrated very promising performance results for a pre-clinical PET scanner based on a single, cooled annulus of monolithic scintillator used with neural networks. Continued development of the system is planned.

Physics in Medicine and BiologyVol. 70(8)
West Virginia University (US), West Virginia University Hospitals (US)
National Institutes of Health, National Institute of Biomedical Imaging and Bioengineering
Openalex Percentile: Top 22%
Medical Imaging Techniques and Applications
1.36
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.