The Verification Process of a <scp>POC</scp> Blood Gas Analyser—The Nova Stat Profile Primer Plus Analyser
The Nova Stat Profile Prime Plus analyser meets the manufacturer's precision and bias claims. Linearity was confirmed for most analytes. There was a good correlation between the Nova and GEM Blood Gas analyser at concentrations within the reference range intervals for all investigated parameters.