Evaluation of Stapes Image Quality with Ultra-High-Resolution CT in Comparison with Conebeam CT and High-Resolution CT in Cadaveric Heads
UHR-CT with and without DLR offers comparable objective image analysis to CB-CT while providing superior subjective image quality. However, this is achieved at the cost of a higher radiation dose. Both CB-CT and UHR-CT with and without DLR are more effective than HR-CT in objective and subjective image analysis.
Authors
- Romain Gillet (ORCID: https://orcid.org/0000-0002-1350-1694)
- Marie Doyen
- Frédérique Groubatch
- Gabriela Hossu (ORCID: https://orcid.org/0000-0003-1178-9097)
- Michaël Eliezer (ORCID: https://orcid.org/0000-0002-4973-1790)
- Ulysse Puel (ORCID: https://orcid.org/0000-0003-2773-3514)
- Alain Blum (ORCID: https://orcid.org/0000-0002-9980-7052)
- F. Boubaker
- Pedro Augusto Gondim Teixeira (ORCID: https://orcid.org/0000-0003-3485-9385)
- Sara Boukhzer
- Cécile Parietti‐Winkler
Institutions
- Inserm (FR)
- ParisTech (FR)
- Centre Hospitalier Régional et Universitaire de Nancy (FR)
- Centre d’Investigation Clinique Innovation Technologique de Nancy (FR)
- Imagerie Adaptative Diagnostique et Interventionnelle (FR)
- Université de Lorraine (FR)
Publication Details
- Journal
- American Journal of Neuroradiology
- Published
- 2025-03-24
- DOI
- https://doi.org/10.3174/ajnr.a8748
- Primary Topic
- Voice and Speech Disorders
- Type
- article
- Field-Weighted Citation Impact
- 0.00